13: Configuration
Table 31
Parameter
Diastolic High Limit Selects the high limit alarm value when
Diastolic Low Limit Selects the low limit alarm value when
Mean High Limit
Mean Low Limit
Table 32
Parameter
Color
SpO
High limit
2
SpO
Low Limit
2
SpO
Desat Limit
2
Table 33
Parameter
Wave 1
142
NBP Settings (Continued)
Description
diastolic is the selected alarm source.
diastolic is the selected alarm source.
Selects the high limit alarm value when
mean is the selected alarm source.
Selects the low limit alarm value when
mean is the selected alarm source.
SpO
Settings
2
Description
Selects the SpO
2
Selects the high alarm limit value.
Selects the low alarm limit value.
Selects the extreme low limit alarm value.
Wave Settings
Description
Selects the waveform displayed in Wave
Sector 1.
Adult: 25-200, 90 (mmHg);
3-27, 12 (kPa)
Infant/Child: 25-120, 70 (mmHg);
3-16, 9 (kPa)
Adult: 20-195, 50 (mmHg);
2-26, 7 (kPa)
Infant/Child: 20-115, 40 (mmHg);
2-15, 5 (kPa)
Adult: 30-220, 110 (mmHg);
4-30, 15 (kPa)
Infant/Child: 30-135, 90 (mmHg);
4-18, 12 (kPa)
Adult: 25-215, 60 (mmHg);
3-29, 8 (kPa)
Infant/Child: 25-130, 50 (mmHg);
3-17, 7 (kPa)
color.
Red, Yellow, Blue, Green, Cyan,
Magenta, White
Adult: 51-100, 100 %
Infant/Child: 51-100, 100 %
Adult: 50-99, 90 %
Infant/Child: 50-99, 90 %
Adult: 50-SpO
Infant/Child: 30-SpO
80 %
Pads, I, II, III, aVR, aVL, aVF, V
Overview
Setting Choices
Setting Choices
Low Limit, 80 %
2
Low Limit
2
Setting Choices