Overview
Table 30
Parameter
NOTE: If the ECG source is pads or paddles, 0.15-40 Hz is used regardless of the
ECG Electrode
Labels
HR/Pulse High
Limit
HR/Pulse Low Limit Selects the default Low Alarm Limit for
VTach HR Limit
VTach Run Limit
PVC Rate Limit
Table 31
Parameter
Color
Unit
NBP Frequency
NBP Alarm Source
Systolic High Limit
Systolic Low Limit
Heart Rate/ECG Settings (Continued)
Description
configuration settings.
Selects the electrode label format. AAMI:
RA, LA, LL, RL, V; IEC: R, L, F, N, C.
Selects the default High Alarm Limit for
the HR derived from the ECG and the
pulse derived from SpO
the HR derived from the ECG and the
pulse derived from SpO
Selects the VTach heart rate limit.
Selects the VTach run limit.
Selects the PVC limit.
NBP Settings
Description
Selects the NBP color.
Selects the measurements units.
Selects the frequency for NBP
measurement: manual or automatic on a
selected schedule.
Selects the alarm source.
Selects the high limit alarm value when
systolic is the selected alarm source.
Selects the low limit alarm value when
systolic is the selected alarm source.
AAMI, IEC
Adult: 35-300, 120 bpm
Infant/Child: 35-300, 160 bpm
.
adjusted in increments of 5
2
Adult: 30-295, 50 bpm
Infant/Child: 30-295, 80 bpm
.
adjusted in increments of 5
2
Adult: 95-150, 100 bpm
Infant/Child: 95-150, 120 bpm
adjusted in increments of 5
Adult: 3-20, 5
Infant/Child: 3-20, 5
Adult: 1-99, 10
Infant/Child: 1-99, 5
Red, Yellow, Blue, Green, Cyan,
Magenta, White
mmHg, kPa
Manual, q1, q2.5, q5, q10, q15, q30,
q60, q120
Systolic, Diastolic, Mean
Adult: 45-260, 160 (mmHg);
6-35, 21 (kPa)
Infant/Child: 45-160, 120 (mmHg);
6-22, 16 (kPa)
Adult: 40-255, 90 (mmHg);
5-34, 12 (kPa)
Infant/Child: 40-155, 70 (mmHg);
5-21, 9 (kPa)
13: Configuration
Setting Choices
Setting Choices
141