MM4006
1
2
3
OFF
ON
ON
2.5.4
RS-485 Point-to-Point Communication Setup
This is the mode for single computer to a single MM4006 controller in long
distance or noisy environment high speed communication.
Front panel DIP switches setting
4
5
6
7
ON
ON
ON
ON
OFF
8
9
10
11
ON
ON
ON
OFF
OFF
ON
OFF
OFF
ON
ON
OFF
ON
ON
OFF
OFF
OFF
ON
ON
OFF
ON
ON
OFF
OFF
OFF
Default Parameters
Parity: NONE
Stop bits: 1
Word length: 8 bits
2.51
12
Description
RS-485 Communication
Address must be Zero
Disable XON/XOFF Mode
Enable XON/XOFF Mode
Terminator: LF
Terminator: CR/LF
Terminator: LF/CR
Terminator: CR
ON
9600 baud
ON
1200 baud
ON
2400 baud
ON
4800 baud
OFF
19200 baud
OFF
38400 baud
OFF
57600 baud
OFF
115200 baud
EDH0181En1020 — 03/03
Local Mode