2 Diagnostic Display Indicators And Controls - DENTSPLY Cavitron JET Plus Installation And Service Manual

Ultrasonic scaler & air polishing prophylaxis system
Hide thumbs Also See for Cavitron JET Plus:
Table of Contents

Advertisement

8 2 Diagnostic Display indicators and Control
Service Indicator
Illuminates when the system
is not functioning properly.
This display has three
distinct modes.
• A fast blink (3 blinks per
second) indicates an
improper set-up.
• A slow blink (1 blink per
second) means the
system is operating out
of factory specifications.
• A steady light indicates
the system is overheating.
Refer to Section 11.1 for
Troubleshooting guidelines.
Power Indicator
Illuminates (3 sec. delay)
when the Main Power ON/
OFF Control Switch is ON
("I" position).
Low Battery
Indicator
Illuminates when
the Tap-On
Foot
TM
Pedal battery power
is approaching end of
life. Replace batteries
as instructed in
Section 7.9.
Purge Control
Illuminates when the Purge function is activated.
To activate Purge, remove insert from the
handpiece, turn the handpiece lavage control
to maximum water flow and press the Purge
button. Water will purge through system for
2 minutes. To deactivate mode during the 2
minute cycle, press Purge button again or press
Tap-On
The Purge Control is also used during the Tap-
On
Foot Pedal Synchronization process. See
TM
Section 7.10 for more details.
13
Boost Indicator
Illuminates when the
Boost Mode has
been activated with
the Tap-On
TM
Turbo Mode
When pressed, extra
power is delivered to the
system, up to 25%. This
will increase the stroke
of the tip. Turbo power
remains on until the
button is pressed again
or unit is turned off.
(Purple arrows illuminate
when Turbo Mode is on).
Foot Pedal.
TM
Foot Pedal.

Hide quick links:

Advertisement

Table of Contents
loading