Touchpoint Plus
Technical Handbook
Maintenance
6) View the reading until it is stable, and then touch 'Apply'.
7) Confirm that the gas reading is stable, and touch 'Next'.
8) Enter the concentration of the span gas and touch 'Start':
9)
When ready touch 'Start' and apply gas to the sensor.
10) View the reading until it is stable, and then touch 'Apply'. This may take 3 – 5 minutes, depending on the
gas type.
11) Confirm that the gas reading is correct.
12) Remove the span gas from the sensor, and confirm that the gas reading has returned to zero.
73