Wire Security Seal (Seal Conversion Kit PN 46-00359)*
1. Temporarily remove the platter and place it in a safe location.
2. Install the calibration button sealing cover and through-hole sealing screw.
3. Tighten the screw and route the wire seal through the holes in the plate and screw.
Figure 59. Calibration Switch/Button Wire Security Seal Assembly
4. Leaving the least amount of slack in the wire loop as possible, close the wire seal and verify the seal
connection is secure.
5. Reinstall the platter.
Figure 60. Platter Installation
* These procedures are for MS2420/MS2430 Scanner/Diva Scale models only.