Download  Print this page

Thermal Array Printer; Noninvasive Pacing - Philips HeartStart MRx Instructions For Use Manual

Hide thumbs

Advertisement

Specifications

Thermal Array Printer

Continuous ECG Strip: The Print key starts and stops the strip. The printer can be configured to run
real time or with a 10-second delay. The strip prints the primary ECG lead and a second or third wave
(75mm printer only) with event annotations and measurements.
Auto Printing: The printer can be configured to automatically print on Mark Events, Charge, Shock,
and Alarm. When an alarm condition occurs, the unit prints the Primary ECG wave, the alarming
wave, if configured, and a third wave (75mm printer only).
Reports: The following can be printed:
• Event Summary (short, medium, and long)
• Vital Sign Trends
• 12-Lead
• Operational Check
• Configuration
• Status Log
• Device Information
Speed: 25 or 50 mm/s with an accuracy of +5%
Amplitude Accuracy: +5% or + 40 uV, whichever is greater
Paper Size:
• 50 mm (W) x 30 m (100 ft.) (L)
• 75 mm (W) x 30 m (100 ft.) (L)

Noninvasive Pacing

Waveform: Monophasic Truncated Exponential
Current Pulse Amplitude: 10 mA to 175 mA (5 mA resolution); accuracy 10% or 5 mA, whichever is
greater
Pulse Width: 40 ms with +10% accuracy
Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy + 1.5%
Modes: Demand or Fixed Rate
Refractory Period: 340 msec (30 to 80 ppm); 240 msec (90 to 180 ppm)
21 Specifications and Safety
279

Advertisement

Table of Contents

This manual is also suitable for:

M3535aM3536a

Comments to this Manuals

Symbols: 0
Latest comments: