The HeartStart MRx is configured to populate each wave sector with a predetermined waveform, when
powered on in Monitor, Manual Defib, or Pacer Mode. Figure 7 shows the default Waves
configuration choices for a device containing all monitor options. See "Configuration" on page 181 for
more information.
A dashed line in a wave sector or an empty wave sector indicates that the waveform source is invalid.
Wave sectors may contain a variety of information, as appropriate to the parameter, view, and task.
Figure 7 Default Waves Configuration

Wave Sector 1

Wave Sector 1 will only contain an ECG waveform. This is the waveform used by the arrhythmia,
heart rate derivation, and AED analysis algorithms.
If Pads are configured as the primary ECG source for Wave Sector 1, the ECG patient cable must be
connected to the HeartStart MRx and to the monitoring electrodes on the patient in order to change
the ECG source to a Leads selection.
When monitoring using a 3-lead ECG set, the HeartStart MRx displays 1 ECG lead at a time.
N O TE
10 Feb 2006 11:20
Mark Event
I nops A rea
P atient name
A dult N on -P aced
ECG/HR alarms
80
HR
T emp
36.9
bpm
120
39.0
50
36.0
Primary ECG
ABP
150
75
0
ABP
mm H g
PAP
mm H g
S p O2 % E t CO2
118/77
24/10
SYS
DIA
100
160
16
(95)
(15)
90
0
PAP
30
15
0
CO2
60
30
0
S oftkey
S oftkey
S oftkey
#1
#2
#3
HH:MM:SS
P ulse
C
NBP
mm H g q 120
bpm
120/80
SYS
80
10:40
160
(95)
90
rhythm status
rpm
mm H g
A w RR
38
18
100
50
30
90
30
M ain M enu
V olume
Displayed Waves
Printed Waves
M easurements /A larms
Patient Info
T rends
O ther
High Contrast On
S oftkey
#4
E xit
2 Getting Acquainted
8
17