Calibration

Calibration
The HeartStart MRx allows you to input the calibration factor (see documentation supplied by the
transducer's manufacturer) of a new transducer plugged into the pressure line.

Known Calibration Factor

To set a known calibration factor, perform the following:
Zero the transducer (see previous section)
1
Press the Menu Select
2
Using the Navigation buttons, select Measurements/Alarms and press the Menu Select button.
3
Select the correct pressure (channel: label) and press the Menu Select button.
4
5
Select Calibration Factor and press the Menu Select button.
Set the calibration factor using the Navigation buttons. Values range from 180.0 to 220.0 in 0.2
6
increments. Press the Menu Select button.
N O TE
Calibration only needs to be done on reusable transducers.
button.