Installation

CAUTION
!
If car adapter X-dock is used as power supply for the
maintenance station only 5 modules at the most may
be connected to the master. You might damage the car
adapter X-dock if you connect additional modules.
The following gas measuring devices can be used with the X-
dock and its modules:
X-dock 5300/6300/6600
with Pac module:
Dräger Pac 3500
Dräger Pac 5500
Dräger Pac 7000
2.4

Limitations on use

Dräger X-dock 5300/6300/6600 and their modules were not
designed to meet the directives for firedamp and explosion
protection and must not be used below ground or in hazardous
locations.
2.5
GPL (General Public Licence)
Some of the device software includes open-source software,
which has been published under GPL, LGPL or another open-
source licence. They include GPL GPLv2, LGPL, MIT,
PostgreSQL, Apache, Apache 2, zlib. The source texts of the
software can be obtained from Dräger on a CD under material
number 83 21 874 for at least three years after purchase of the
software. The licence terms and conditions of the software are
included on CD.
3
Installation
WARNING
!
Risk of personal injury and damage to equipment
through
faulty
instruments.
Unless the maintenance station is properly set up for
the scheduled maintenance tasks, there is a risk that
gas detection instruments are not properly serviced.
Prior to initial operation must be approved by a
qualified specialist testing the maintenance station
with appropriately configured instruments for correct
performance of the specified tasks.
The maintenance station can automatically recognise the test
gases required for the instrument and cross-check them with
the connected and configured test gases.
The gas exposure sequence always results from the sequence
of the connected test gas cylinders.
The maintenance station has various safety mechanisms at its
disposal
to
prevent
configurations
For example there are limitations with regard to specific test
gas concentrations or an automatic flushing process takes
place at the beginning of the test in case of high measured
values. Nevertheless, it is necessary for a qualified specialist
to design and approve the station for the required task.
18
with X-am 125 (+) module:
Dräger X-am 1700
Dräger X-am 2000
Dräger X-am 2500
Dräger PARAT 5000
Dräger PARAT 5600*)
maintenance
of
gas
critical
to
The designer for example must take into consideration the
cross-sensitivities of the sensors with connected test gases
and consult the respective sensor data sheets. A description of
the tasks to be accomplished must be provided which in turn
determines which test method with which test gas
concentration is appropriate.
If the required expert knowledge is not available, it has to be
obtained from others (e. g. specialists, test institutions or
manufacturers).
NOTICE
i
i
Ensure adequate space for the entire assembly.
The master and all modules must have the same
firmware version. If this is not the case, a firmware
update needs to be carried out (see chapter 6.4 on
page 25).
1. If applicable, install modules on the master station in
accordance with the assembly instructions (X-dock 6300/
6600 only).
A maximum of 10 modules can be installed on one
master station.
The available modules can be combined in any way
desired.
2. If applicable, install wall mount or cylinder holder in
accordance with the assembly instructions.
3. Remove the sleeves from the intended gas inlets and from
the gas outlet.
NOTICE
i
i
If the sleeve is not removed from the gas outlet, the
station will be unable to conduct the self-test without
errors.
4. Fit the gas feed hoses to the gas inlets on the master and
connect them to the regulator valve on the test gas
cylinder.
NOTICE
i
i
detection
For the gas exposure sequence Dräger recommends
using toxic gases with increasing concentrations.
Dräger recommends not to exceed a hose length of
10 m for the gas supply hoses.
5. If required, connect an exhaust hose (max. 10 m long) to
the exhaust outlet.
6. Ensure a supply of compressed air or fresh air:
Connect the compressed air hose to the compressed
air port (outlet pressure of pressure regulator valve
0.5 bar, flow rate >3 l/min).
Adjust fresh air input (see chapter 4.6.1 on page 21).
Setting: Supplied by compressed air input
OR
If required, connect the fresh air hose to the fresh air
safety.
filter.
Adjust fresh air input if necessary (see chapter 4.6.1 on
page 21). Setting: Supplied by pump.