3
Additional Information
Cleaning
Requirements
The Storage
and Shipment
Environment
HP
16500A/16501A
Additional Information
This chapter contains additional information about maintaining the HP
16500N1650lA. You learn how to clean the instrument and how to
store, ship, and tag it for service or repair.
This
chapter also provides
direction for where you should turn to next to learn how to use the HP
16500N1650lA.
When cleaning the HP 16500A or the HP 165OlA, use mild soap and
water only. A harsh soap or solvent may damage the water-base paint
ftnish.
Clean the CRT display and surrounding area regularly. Do not place
tape or other foreign material on the screen that may interfere with the
operation of the touchscreen.
Vacuum the ventilation slots on the sides of the instrument and the fan
on the rear panel whenever there is a visible amount of dust on them.
The instrument may be stored or shipped in environments within these
limits:
Temperature: -40 to 70 degrees C (-40 to 158 degrees
F)
Humidity: Up to 90% relative humidity at 65 degrees C
(149 degrees F)
Altitude: Up to 15,300 m
(50,000 ft)
The instrument should also be protected from temperature extremes,
which cause condensation within the instrument. Condensation within
the instrument may cause the instrument to malfunction when it is
operating.