Force Contact Outputs - GE T35 Instruction Manual

Transformer management relay ur series
Hide thumbs Also See for T35:
Table of Contents

Advertisement

5.9 TESTING
!"
PATH: SETTINGS
TESTING
# FORCE CONTACT
# OUTPUTS
The relay contact outputs can be pre-programmed to respond to the Test Mode.
If set to "Disabled", the contact output remains fully operational. If operates when its control operand is Logic 1 and will
resets when its control operand is Logic 0. If set to "Energize", the output will close and remain closed for the entire duration
of the Test Mode, regardless of the status of the operand configured to control the output contact. If set to "De-energize",
the output will open and remain opened for the entire duration of the Test Mode regardless of the status of the operand con-
figured to control the output contact. If set to "Freeze", the output retains its position from before entering the Test Mode,
regardless of the status of the operand configured to control the output contact.
These settings are applied two ways. First, external circuits may be tested by energizing or de-energizing contacts. Sec-
ond, by controlling the output contact state, relay logic may be tested and undesirable effects on external circuits avoided.
Example 1: Initiating a Test from User-Programmable Pushbutton 1
5
The Test Mode should be initiated from User-Programmable Pushbutton 1. The pushbutton will be programmed as
"Latched" (pushbutton pressed to initiate the test, and pressed again to terminate the test). During the test, Digital Input 1
should remain operational, Digital Inputs 2 and 3 should open, and Digital Input 4 should close. Also, Contact Output 1
should freeze, Contact Output 2 should open, Contact Output 3 should close, and Contact Output 4 should remain fully
operational. The required settings are shown below.
To enable User-Programmable Pushbutton 1 to initiate the Test mode, make the following changes in the
!
TESTING
TEST MODE
TEST MODE FUNCTION:
Make the following changes to configure the Contact I/Os. In the
FORCE CONTACT INPUTS
"Disabled",
FORCE Cont Ip 1:
"Freeze",
FORCE Cont Op 1:
Example 2: Initiating a Test from User-Programmable Pushbutton 1 or through Remote Input 1
The Test should be initiated locally from User-Programmable Pushbutton 1 or remotely through Remote Input 1. Both the
pushbutton and the remote input will be programmed as "Latched". The required settings are shown below.
Write the following FlexLogic™ equation (enerVista URPC Setup example shown):
Set the User Programmable Pushbutton as latching by changing
!
PUSHBUTTONS
USER PUSHBUTTON 1
Input 1 to initiate the Test mode, make the following changes in the
TEST MODE FUNCTION:
5-114
!"

FORCE CONTACT OUTPUTS

FORCE Cont Op 1
:Disabled
FORCE Cont Op 2
MESSAGE
:Disabled
FORCE Cont Op xx
MESSAGE
:Disabled
menu:
"Enabled" and
TEST MODE INITIATE:
menus, set:
"Open",
FORCE Cont Ip 2:
"De-energized",
FORCE Cont Op 2:
!
PUSHBUTTON 1 FUNCTION
"Enabled" and
TEST MODE INITIATE:
T35 Transformer Management Relay
Range: Disabled, Energized, De-energized, Freeze
Range: Disabled, Energized, De-energized, Freeze
Range: Disabled, Energized, De-energized, Freeze
"
"
PUSHBUTTON 1 ON
!"
SETTINGS
TESTING
"Open", and
FORCE Cont Ip 3:
"Open", and
FORCE Cont Op 3:
!
SETTINGS
PRODUCT SETUP
to "Latched". To enable either Pushbutton 1 or Remote
!"
SETTINGS
TESTING
"
"
VO1
5 SETTINGS
5.9.3 FORCE CONTACT OUTPUTS
SETTINGS
!"
FORCE CONTACT INPUTS
"Closed"
FORCE Cont Ip 4:
"Disabled"
FORCE Cont Op 4:
!"
USER-PROGRAMMABLE
!
menu:
TEST MODE
GE Multilin
!"
and