   Summary of Contents for Philips EverFlo

  • Page 1 EverFlo EverFlo Q 1053621 R05 LZ 02/09/2015 EN-INTL, DE...
  • Page 2: Table Of Contents

    Symbol Key ....................... 2 Abbreviations ....................2 Chapter 1: Introduction ..............3 Intended Use ....................3 About Your EverFlo / EverFlo Q ..............3 Parts of Your Concentrator ................3 Accessory Equipment and Replacement Parts ........4 Warnings and Cautions ................4 Chapter 2: Operating Instructions..........
  • Page 3: Symbol Key

    Compliant with the Waste Electrical and Electronic Equipment/Restriction of the Use of European Declaration of Certain Hazardous Substances in Electrical and Conformity Electronic Equipment (WEEE/RoHS) recycling directives Abbreviations Light Emitting Diode Liters per Minute Oxygen Percentage Indicator © 2014 Koninklijke Philips N.V. All rights reserved.
  • Page 4: Chapter 1: Introduction

    Intended Use The EverFlo / EverFlo Q Oxygen Concentrator is intended to provide supplemental oxygen to persons requiring oxygen therapy. The device is not intended to be life supporting or life sustaining. About Your EverFlo / EverFlo Q The device produces concentrated oxygen from room air for delivery to a patient requiring low flow oxygen therapy.
  • Page 5: Accessory Equipment And Replacement Parts

    Note: When connected to the UltraFill Oxygen Filling Station, the EverFlo device disables its Low Flow Alarm when a cylinder is in the process of being filled. If you are breathing from the EverFlo at this time and an occlusion occurs in the cannula, the EverFlo will not sense the disruption of the flow of oxygen to you. If you are breathing from the EverFlo device while filling a cylinder, ensure that you place your oxygen cannula in a position to avoid it being kinked or crushed.
  • Page 6 • Application accessories shall include a means to reduce the propagation of fire. • The use of accessories, transducers, and cables other than those specified by Philips Respironics may result in increased emissions or decreased immunity of the device. •...
  • Page 7: Chapter 2: Operating Instructions

    Step 3-B4 Fill your humidifier bottle according to the manufacturer’s instructions. Mount the filled humidifier on the top of the EverFlo / EverFlo Q device inside the Velcro strap, as shown in the illustration on the right. Tighten the Velcro strap around the bottle and secure it so it is held firmly in place.
  • Page 8 EverFlo / EverFlo Q User Manual Press the power switch to the On [I] position. Initially, all the LEDs will illuminate and the audible alert will beep for a few seconds. After that time, only the green LED should remain lit. You can begin breathing from the device immediately even though it typically takes 10 minutes to reach oxygen purity specifications.
  • Page 9: Chapter 3: Cleaning & Maintenance

    Caution: Excess moisture may impair the proper operation of the device. Cleaning Periodically, use a damp cloth to wipe down the exterior case of the EverFlo / EverFlo Q device. If you use medical disinfectants, be sure to follow the manufacturer’s instructions.
  • Page 10: Chapter 4: Alarms And Troubleshooting

    EverFlo / EverFlo Q User Manual Chapter 4: Alarms and Troubleshooting Alarm and Indicators The device has an audible alarm and three LED indicators, as shown below. NORMAL Green Power On Indicator Yellow Check System Indicator Red Alarm Indicator Audible Alarm / Colored LED...
  • Page 11: Troubleshooting Guide

    EverFlo / EverFlo Q User Manual Troubleshooting Guide Problem Why it Happened What to Do Turn the flow rate down to your prescribed level. Wait at least 2 minutes. Yellow LED is blinking. The Red LED is off and the...
  • Page 12: Chapter 5: Specifications

    EverFlo / EverFlo Q User Manual Chapter 5: Specifications Environmental Operating Transport & Storage Temperature 13 to 32° C -34 to 71° C Relative Humidity 15 to 95%, noncondensing 15 to 95%, noncondensing Altitude 0 to 2286 m Physical Dimensions 58 cm x 38 cm x 24 cm...
  • Page 13 1039366, 1039364 1039365, 1039368 1039370, 1102443 1104000 1020007, 1020014 <40 dBA typical 1020015, 1039367 Classification The EverFlo / EverFlo Q Oxygen Concentrator is classified as: • IEC Class II Equipment • Type BF Applied Part • IPX1 Drip Proof •...
  • Page 14: Appendix A: Emc Information

    EverFlo / EverFlo Q User Manual Appendix A: EMC Information ’ uidance and anufacturer eclaration lectroMaGnetic Missions This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
  • Page 15 EverFlo / EverFlo Q User Manual ’ uidance and anufacturer eclaration lectroMaGnetic MMunity This device is intended for use in the electromagnetic environment specified below. The user of this device should assure that it is used in such an environment.
  • Page 16: Limited Warranty

    (3) years from the date of sale by Respironics, Inc. to the dealer. Respironics warrants that the EverFlo / EverFlo Q units serviced by Respironics, or an authorized service center, will be free from defects in serviced materials for a period of 90 days and free from defects in workmanship for a period of 90 days from the time of service.

This manual is also suitable for:

Everflo q

Comments to this Manuals

Symbols: 0
Latest comments: