B) Connecting The 6 Pin Mini Din Connector - Omega 3200 User Manual

Thermal mass flow controllers
Hide thumbs Also See for 3200:
Table of Contents

Advertisement

b)
Connecting The 6 Pin Mini Din Connector
Using a suitable mating connector the pins of the integrated connector
should be wired as follows:
Connecting To The Integrated 6 Pin Connector
Pin 2 should be connected to the Positive of the power source.
Pin 6 should be connected to the Negative (Ground) of the power source.
Pin 3 provides the signal output and should be connected to the positive
terminal of the display, data acquisition system or voltmeter.
Pin 1 is the signal negative (ground) and should be connected to the
negative (Ground) terminal of the display, data acquisition system or
voltmeter.
Pin 4 provides the input signal and should be connected to the positive
terminal of the voltage source. The (0-5VDC) voltage control signal should
be supplied from a low impedance source.
Pin 5 is the input signal negative (ground) and should be connected to
the negative (Ground) terminal of the voltage source.
Caution: Avoid high voltage static discharges to the input signal
connection. Do not short the input/output signal wires or allow
them to contact the power wires at any time. DAMAGE WILL
RESULT!
Pin Out of Integrated
Connector
M-4271/0707, pg. 9 of 26