6. The procedure will ask for the USB color sensor to be placed on the center of the display panel (Figure S.3). Tilt the display
panel approximately 5˚ backward and place the USB color sensor in the center of the display panel (Figure S.1).
NOTE:
Place the USB color sensor fl at against the LCD to avoid external light contamination. DO NOT press the calibrator
against the display panel.
Press "SELECT" to start the calibration. Stand-alone calibration may take several minutes depending on user settings.
7. After the CALIBRATION SUCCEEDED message appears (Figure S.4), press SELECT.
8. To end the calibration mode, press "EXIT".
Figure S.3
Figure S.4
English-35