Menu
Auto Adjust
Target Color Temp
x
y
Target Luminance
Probe
Function and operation ([
Starts adjustment.
Note
This menu is not selectable in the following case:
• When 4:4:4 XYZ 12bit or DVI Computer is selected in the Format menu of the Input
Configuration menu
Before automatic adjustment, select the value to use for the adjustment in following
sequence.
When you use the previously adjusted values for adjustment, select Auto Adjust to start
adjustment without performing step 1, step 2 and step 3.
1 Select the color temperature to be used in Target Color Temp.
When Any is selected, enter the values of the CIE color system x and y coordinates.
2 When Contrast Hold is set to Off, select Target Luminance and enter the value of the
2
luminance (cd/m
) when the 100% white signal is input.
3 Select the probe to be used in Probe.
4 Select Auto Adjust.
The following message appears.
Set probe on cursor.
Attach the probe on the center of the display to start automatic adjustment. (It may take
some time until the automatic adjustment is completed.)
Note
First select the probe and then select Auto Adjust.
To cancel/stop the adjustment
Press the MENU button.
Sets the adjustment value to be used from D93, D65, D61, D56, D-Cine or Any.
Enters the CIE color system x coordinate.
Set to 0.280 to 0.345.
Enters the CIE color system y coordinate.
Set to 0.280 to 0.352.
2
Enters the luminance (cd/m
) when the 100% white signal is input.
Adjust the luminance by the contrast adjustment (backlight adjustment).
Set to 40 to 250. [100]
Notes
• This menu is not selectable when Contrast Hold is set to On.
• The luminance is reduced to about half due to the principle of operation in the black frame
insertion mode or the interlace display mode. However, on this monitor the normal
luminance is set by the contrast adjustment (backlight adjustment). For this reason, as the
upper limit of Target Luminance becomes about 120 [cd/m
• The luminance calibrated internally by the backlight does not guarantee an absolute value
and may be different from the luminance calibrated by the probe. When you set Target
Luminance near to the upper limit (250) or the lower limit (40), the setting value may
exceed the adjustment range and cause an error. If an adjustment error occurs, change
Target Luminance and perform the automatic adjustment again.
Selects the probe to be used for automatic adjustment of the color temperature.
[CA-210]: Select when Konica Minolta CA-210 is used.
PM 5639/06: Select when DK-Technologies PM 5639/06 is used.
Eye-One Pro: Select when X-Rite Eye-One Pro is used.
The cable to connect the probe and the monitor is required.
About the color temperature after automatic adjustment
The values of x and y displayed by the probe may be different from those set in Target Color
Temp because the x and y values of the probe is calibrated by the monitor.
Note
An external PC with USB connector and software for adjusting the color temperature are
required to use the Eye-One Pro. Before adjusting the color temperature automatically, start
the software for adjusting the color temperature.
For the software for adjusting the color temperature, please contact your Sony representative.
For more information about the cable specification required and about the connection, see
"Connection Cable Specifications for Color Temperature Probes" on page 123.
]: factory setting)
2
], set it to less than 120 [cd/m
2
].
47
Adjustment Menu