Product Information and User Instructions Models CH5000S and CH5000 Caution – This device is only to be used by or on the prescription of a physician.
Warranty Page 11 Intended Use The intended use of the Pure Oxygen Concentrator is to provide supplemental oxygen to those individuals needing supplemental oxygen therapy. The device is not intended to be life supporting or life sustaining. Introduction to Your Pure Oxygen Concentrator The Pure Oxygen Concentrator is designed as a stationary supplemental oxygen source.
Operation Normal, Power On • Red Operation Alarm, Service Required NOTE: For both the CH5000S and the CH5000 models, the Green LED will blink for 6-10 minutes during start up then become solid Green during normal operation. Power Switch: I/0 •...
(page 9) may result in decreased unit performance and oxygen purity levels. NOTE: The use of certain humidifiers and administration of accessories not specified for use with this oxygen concentrator may impair its performance........................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000 ........................
8 if there is no oxygen flow. 12) Put on the cannula as directed by your equipment provider. 13) When not in use, turn off the concentrator. Figure F Figure G ........................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000 ........................
HEPA Intake Filter 1/each PL-2001854 Reusable Foam Intake Filter 1/each PL-3000706 Filter Intake Door 1/each The following humidifier bottle is recommended for use with the Pure Oxygen Concentrator. Accessories Description HUM 001 Humidifier Bottle 50/case HUM-TUB Humidifier Adapter Tubing 50/case...
Alarms and Troubleshooting Alarm Table – Oxygen Sensor Model Number CH5000S Light Indication Sounds Cause Blinking Green None Normal Operation - Unit in Start-up Phase Solid Green None Normal Operation Solid Yellow None Oxygen Purity Between 72% and 81%. Device can still be used but service provider should be contacted.
Page 8
Turn your unit “OFF”, switch to your reserve occur with your oxygen / alternate oxygen system, and contact your concentrator or none of service provider immediately. the solutions above work ........................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000 ........................
16.75" W x 10.25" D x 24.5" H Weight 35lbs Cabinet UL94V-O Flame Retardant Rated Thermoplastic Electrical Double Insulated Cord - Do Not Use an Extension Cord Duty Cycle Continuous Use Warranty Limited 3 Year ........................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000 ........................
RF Communications equipment and the Pure Oxygen Concentrator The Pure Oxygen Concentrator is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the Pure Oxygen Concentrator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications (transmitters) and the Pure Oxygen Concentrator as recommended below.
Warranty The Pure Oxygen Concentrator has been carefully manufactured and inspected in our Lehigh Acres, FL USA facility and is warranted to be free from defects in workmanship and materials. Under this warranty, Drive Medical and its subsidiaries obligation shall be limited to the replacement or repair of any such units or parts that prove, by Drive Medical’s...
Page 12
Drive Medical 99 Seaview Blvd Port Washington, NY 11050 Toll-free 877-224-0946 www.drivemedical.com...