Diode Test

Test Current ............................................................ 100 μA or 1 mA
Response Time....................................................... 300 samples/sec with audible tone.
Accuracy is given as ± (% measurements + % of range)
24 Hour
Range
(23 ±1 °C)
5.0000 V
0.002 + 0.002
10.0000 V
0.002 + 0.001

Measurement Rates (IEEE488[4])

Function
DC Volts, DC Current, and
Resistance
[2]
AC Voltage and AC Current
Frequency and Period
Capacitance
[1] Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off.
[2] Maximum measurement rates for 0.01 % of ac step. When dc input varies, additional settling delay is required.
[3] For remote operation or external trigger using default settling delay
[4] Speeds available in OutG SW 1.0.700.18 or higher. Note that the measurements rates for RS232 can vary depending on the
baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurement/s. The LAN bus
has a maximum measurement rate of 963 measurement/s.
90 Days
(23 ±5 °C)
0.008 + 0.002
0.01 + 0.002
0.008 + 0.002
0.01 + 0.002
Integration Time
Digits
Setting
60 Hz (50 Hz)
100 NPLC
10 NPLC
167 (200) ms
1 NPLC
16.7 (20) ms
0.2 NPLC
0.02 NPLC
3 Hz
20 Hz
[3]
200 Hz
1 s
100 ms
10 ms
Introduction and Specifications
Temperature
1 Year
Coefficient/ °C
(23 ±5 °C)
Outside 18 to 28 °C
0.001 + 0.002
0.001 + 0.002
Measurements/Second
8845A
1.67 (2) s
0.6 (0.5)
6 (5)
60 (50)
3.3 ms
270
500 us
995
0.47
1.64
4.5
1
9.8
80
NA
1
Electrical Specifications
[1]
8846A
0.6 (0.5)
6 (5)
60 (50)
270
995
0.47
1.64
4.5
1
9.8
80
2
1-23

Hide quick links: