Calibration Without A Pc - Canon DP-V2410 Instruction Manual

Table of Contents

Advertisement

Calibration without a PC (

When "User 1-7" under "Picture Mode" is selected, you can perform calibration using an external sensor, without using
the computer.
The supported external sensors are Konika Minolta Display Color Analyzers CA-310 and CA-210. Be sure to also read
the instruction manual of the CA-310 and CA-210.
1. Connect the display color analyzer to the USB port of the main unit.
2. Open the OSD menu and select "Adjustment" → "Detail Settings" → "Calibration".
Set each target value.
3. Press the jog dial and select "Start".
Please follow the information indicated on the screen.
4. Initialize the sensor.
Set the mode dial of the Universal Measuring Probe to "0-CAL".
Press the jog dial of the video display, select "OK", and execute initialization.
5. Place the universal measuring probe pointing at the center of the video display.
Set the mode dial of the universal measuring probe to "MEAS" and place the probe as shown below according to
the displayed content. Press the jog dial of the video display, select "OK", and execute calibration.
6. Finish calibration.
When the message "Calibration is completed". is displayed, press the jog dial and select "OK".
• If the message "Calibration error." is displayed.
Calibration has been terminated due to an error. The main unit returns to the state before calibration. (
• To cancel calibration
Press the jog dial during calibration and select "Cancel". The main unit returns to the state before calibration.
22
Operating the Video Display
40)
Main Unit
Konica Minolta
Display Color Analyzer CA-310, CA-210
Konica Minolta Universal Measuring Probe
CA-310 support: CA-PU32, CA-PU35
CA-210 support: CA-PU12, CA-PU15
83)