Wrist Blood Pressure Monitor
Model R2
• Instruction Manual
• Mode d'emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
IM-HEM-6113-E(V)-02-10/2011
EN
FR
DE
IT
ES
NL
RU
AR
5330293-3B

Advertisement

Table of Contents
loading

Summary of Contents for Omron R2

  • Page 1 Wrist Blood Pressure Monitor Model R2 • Instruction Manual • Mode d’emploi • Gebrauchsanweisung • Manuale di istruzioni • Manual de instrucciones • Gebruiksaanwijzing • IM-HEM-6113-E(V)-02-10/2011 5330293-3B...
  • Page 2: Table Of Contents

    Contents Introduction ............. 3 Important Safety Information ......4 1. Overview ............. 6 2. Preparation ..........9 2.1 Installing/Replacing the Batteries .... 9 3. Using the Unit ........... 11 3.1 Applying the Wrist Cuff ......11 3.2 Correct Posture ........13 3.3 Taking a Reading ........
  • Page 3: Introduction

    Introduction Thank you for purchasing the OMRON R2 Wrist Blood Pressure Monitor. This remarkable, compact and easy to use instrument is ideal for people who frequently monitor their own blood pressure. The small, pre-formed wrist cuff is very convenient and easy to apply.
  • Page 4: Important Safety Information

    Important Safety Information Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis. Please read this section carefully before using the unit. Warning: • Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) •...
  • Page 5 Important Safety Information (Battery Usage) • If battery fluid should get on your skin or clothing, immediately rinse with plenty of clean water. • Use only two “AAA” alkaline batteries with this unit. Do not use other types of batteries. •...
  • Page 6: Overview

    1. Overview Main Unit A. Display C. Wrist cuff B. Battery compartment D. O/I START button cover E. MEM (Memory) button...
  • Page 7: Display

    1.Overview Display F. Systolic blood pressure J. Heartbeat symbol G. Diastolic blood pressure 1. Flashes during measurement H. Memory symbol 2. If flashing after (Displayed when viewing measurement completed or values stored in memory) when viewing results Battery low symbol stored in the memory, indicates blood pressure out of recommended...
  • Page 8: Package Contents

    1.Overview Package contents M. Two “AAA” alkaline (LR03) N. Storage case batteries • Blood pressure pass • Instruction manual • Blood Pressure Monitor unit • Guarantee card...
  • Page 9: Preparation

    2. Preparation Installing/Replacing the Batteries Remove the battery compartment cover by pulling it off in the direction of the arrow. 1) Push the hook on the bottom of the battery cover. 2) Pull the cover off the main unit. Insert two identical 1.5V “AAA” alkaline (LR03) batteries in the battery compartment.
  • Page 10 2.Preparation Put the battery cover back in place. Note: The measurement values continue to be stored in memory even after the batteries are replaced. Battery Life & Replacement If the battery low symbol ( ) appears on the display, replace both batteries at the same time. - When the battery low symbol ( ) starts to blink, you will still be able to use the unit for a short while.
  • Page 11: Using The Unit

    The blood pressure can differ between the right arm and the left arm and therefore also the measured blood pressure values can be different. Omron recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your physician which arm to use for your measurement.
  • Page 12 3.Using the Unit Hold the bottom part of the wrist cuff and wrap it around the wrist while pulling so that it fits comfortably. Notes: • Unless the wrist cuff is wrapped securely around the wrist, it will not be possible to take correct measurements.
  • Page 13: Correct Posture

    3.Using the Unit Correct Posture Correct posture during measurement helps you get accurate results. You should also try to measure your blood pressure at the same time each day. Notes: • Do not take measurement after bathing, drinking alcohol, or exercising. •...
  • Page 14: Taking A Reading

    3.Using the Unit Taking a Reading Press the O/I START button to start measurement. The wrist cuff will start to inflate automatically after the power is turned on. After the unit has detected your blood pressure and pulse rate, the cuff automatically deflates and your blood pressure and pulse rate are displayed.
  • Page 15 3.Using the Unit Check the measurement result. The unit automatically stores blood pressure and pulse rate into its memory. Refer to “3.4 Using the Memory Function”. Undo the cuff and remove the unit. Press the O/I START button to turn off the unit. If you forget to turn off the unit, it will shut itself off automatically after two minutes.
  • Page 16: Using The Memory Function

    3.Using the Unit Using the Memory Function This unit has a memory capable of storing 30 sets of readings. Every time you complete the measurement, the unit automatically stores blood pressure and pulse rate. Note: When 30 sets of readings are stored in memory, the oldest set will be deleted to store a new set.
  • Page 17 3.Using the Unit To Delete All Values in Memory You cannot delete the stored readings partially, all the readings in the unit will be deleted. To delete stored readings, first press the MEM button. Then while holding it down, press the O/I START button simultaneously for about 2-3 seconds.
  • Page 18: Quick Reference Guide

    You may require more rest time between measurements depending on your individual physiological characteristics. • Only use the R2 to measure your own blood pressure to avoid confusion as the results of measurements are stored in memory. • Always wrap the wrist cuff around your wrist before starting to take a measurement.
  • Page 19: Handling Errors And Problems

    Carefully read and repeat the not fastened steps listed under “3.1 securely. Applying the Wrist Cuff”. An E mark with a code/number Consult your OMRON indicates the device has a retail outlet or distributor. hardware failure. Replace the batteries with The battery two new “AAA”...
  • Page 20: Troubleshooting

    (Refer to “3.3 measurement. Taking a Reading”.) Consult your Wrist cuff pressure does Air is leaking from OMRON retail not rise. the wrist cuff. outlet or distributor. Apply the cuff correctly so that it is securely Wrist cuff deflates too...
  • Page 21 (+/-) polarity. • Press the O/I START button and repeat measurement. • If the problem continues, try replacing Other problems. the batteries with new ones. If this still does not solve the problem, contact your OMRON retail outlet or distributor.
  • Page 22: Maintenance And Storage

    6. Maintenance and Storage To protect your unit from damage, please avoid the following: • Subjecting your unit to extreme temperatures, humidity, or direct sunlight. • Washing the cuff or exposing the cuff or unit to water. • Disassembling the unit. •...
  • Page 23 • It is generally recommended to have the unit inspected every two years to ensure correct functioning and accuracy. Please consult your authorised OMRON dealer or the OMRON Customer Service at the address given on the packaging or attached literature.
  • Page 24: Technical Data

    7. Technical Data Product Description Wrist Blood Pressure Monitor OMRON R2 (HEM-6113-E(V) / Model HEM-6113-E8(V)) Display LCD Digital Display Measurement Oscillometric method Pressure: 0 to 299 mmHg Measurement Range Pulse: 40 to 180 beats/min Memory 30 Measurements Pressure: Within ±3 mmHg Accuracy Pulse rate: Within ±5% of reading...
  • Page 25 7.Technical Data • This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the Pressure Sensor, is produced in Japan. • Disposal of this product and used batteries should be carried out in accordance with the national regulations for the disposal of electronic products.
  • Page 26 This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed: Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device.
  • Page 27: Some Useful Information About Blood Pressure

    8. Some Useful Information about Blood Pressure What is Blood Pressure? Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart’s cycle. The highest pressure in the cycle is called the Systolic Blood Pressure;...
  • Page 28 8.Some Useful Information about Blood Pressure Upper curve: systolic blood pressure Lower curve: diastolic blood pressure Time of day Example: fluctuation within a day (male, 35 years old) Classification of Blood Pressure by the World Heath Organization The World Health Organization (WHO) and the International Society of Hypertension (ISH) developed the Blood Pressure Classification shown in this figure.
  • Page 29 Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS www.omron-healthcare.com OMRON HEALTHCARE Production facility MANUFACTURING VIETNAM CO., LTD. Binh Duong Province, VIETNAM OMRON HEALTHCARE UK LTD.
  • Page 30 Notes...