20 Troubleshooting

20.1 Section overview

This section covers common problems encountered when using the RM500SL. If
you require further assistance, note your model and serial number (located on the
recessed panel) and contact your local distributor, or Audioscan at 519-268-3313
800-265-2093 (USA only), 519-268-3256 (fax) or service@audioscan.com
PLEASE DO NOT SHIP YOUR UNIT BACK TO THE FACTORY. In most cases
we can courier you a part that you can quickly and easily replace yourself.
This section covers the following topics:

20.2: Self test failures

20.3: HIT high distortion or noise
20.4: HIT curves inconsistent
20.5: HIT curves differ from specifications
20.6: HIT speaker overdriven
20.7: No HIT reference mic. detected
FastFacts 20.1: Troubleshooting
20.2 Self test failures
Any failure during power-on self test (P.O.S.T.) is indicated on-screen and further
use of the instrument is denied. Try the following steps to resolve the failure:
1) To access Self test, press <Setup>, then highlight and <PICK> [Self test].
2) If you received a Max out/in or a Routing failure, remove any hearing aids from
the test chamber, and rerun self test.
3) Unplug all microphones (REM & HIT). Rerun self test. If self test passes,
reconnect the microphones, one by one, running self test each time until a
failure occurs. If the failure returns, display and print the test results and
contact Audioscan for help.
FastFacts 20.2: Self test failures
0611
20.8: Invalid HIT calibration
20.9: No REM ref. mic. detected
20.10: Invalid REM calibration
20.11: REM speaker overdriven
20.12: Barcode scanner malfunction