Advanced Operations
Program Type
Single Wavelength
Multi Wavelength
Free Programming
62
Program Number:
Specific test number, with which the program can subsequently be
called up from the selection list in the User Programs menu or the
Favorites menu.
2. Use the alphanumeric keypad to enter a program number
3. Press OK.
Program Name:
4. Use the alphanumeric keypad to enter a program name. The
5. Press Back to go back to the previous program point or press
Program Type:
6. Select the required option
7. If the Single Wavelength
Table 8 Program descriptions
Description
Measurements at a defined wavelength
In the Multi Wavelength mode, absorbance values can be measured at up to four wavelengths
and the results can be mathematically processed to obtain sums, differences and relationships.
This is an advanced form of programming for original user-developed methods. In Free
Programming, the user defines the measurement process, variables and calculations involved in
obtaining a reading.
between 950 and 999. The lowest available number appears
automatically.
Note: If the program number is already assigned to another user
program, a message appears, asking whether the existing program
should be replaced. Press OK to overwrite the existing program.
name can be a maximum of 28 characters long.
Next to continue with the input of the program data.
Wavelength
(section 6.1.1.2 on page
unit, wavelength, absorbance formula, wavelength λx,
concentration factor Kx, concentration resolution, chemical
form and calibration equation.
More information on Free Programming parameters see
section 6.1.2 on page
72.
(Table
8) and press Next.
(section 6.1.1.1 on page
64) is selected, define the
63) or Multi