Invacare® 9000 Series Wheelchair 9000 SL User Manual 9000 XT 9000 XDT 9000XT Recliner This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Troubleshooting..............................................35 Maintenance ................................................36 FRONT RIGGINGS .....................................39 Installing/Removing Front Riggings........................................39 Adjusting Footplate Height ..........................................41 Raising/Lowering Elevating Legrest Assembly ....................................43 Installing Impact Guards/Calf Strap........................................44 Replacing Heel Loop ............................................44 Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 4
Installing/Replacing Six or Eight-Inch Front Casters and Forks..............................60 Adjusting Front Caster Mounting........................................61 10 ANTI-TIPPERS/WHEEL LOCKS ..............................62 Installing/Adjusting Anti-tippers........................................62 Using/Adjusting Disk Wheel Locks........................................65 Installing Wheel Lock Shoe Extensions ......................................69 Invacare® 9000 Series Wheelchair Part No. 1056953...
Caution indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor injury or both. IMPORTANT Indicates a hazardous situation that could result in damage to property if it is not avoided. Gives useful tips, recommendations and information for efficient, trouble-free use. Invacare® 9000 Series Wheelchair Part No. 1056953...
Invacare warrants the side frames and cross members of this product when purchased new and unused to be free from defects in materials and workmanship for a period of five (5) years from the date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty.
2 OVERVIEW 2 Overview Label Locations Crossmember Serial Number Label is located on the cross Lower Frame brace on the left side of Tube the chair 9000 XDT Only Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 9
LIFT MECHANISM TO OF AXLE BUSHINGS. ENSURE BE PERFORMED BY THAT PLUNGER PIN IN HEAD END PROPERLY TRAINED AUTHORIZED SERVICE OF AXLE IS FULLY RELEASED. PERSONNEL ONLY. KEEP DETENT BALLS CLEAN. 1022642 1001948 Part No. 1056953 Invacare® 9000 Series Wheelchair...
WEIGHT pkg) Swing Back Removable, Conventional, (APPROX.) Adjustable Height (Adjustable back angle style) FRONT RIGGINGS Swingaway footrests and elevating legrests REAR WHEELS 20, 22 and 24*-inch composite; Urethane*, pneumatic or pneumatic-flat free insert Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 11
Swing back removable, space saver* or WEIGHT LIMIT 250 lbs (114 kg) conventional in desk* or full length and in fixed* or adjustable height FRONT RIGGINGS Swingaway footrests/elevating legrests WHEEL LOCKS Toggle lock -push or pull Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 12
60½ lbs Desk* or full length and in (APPROX.) 18x17-inch seat frame with complete package fixed* or adjustable height FRONT RIGGINGS Swingaway footrests and WHEEL LOCKS Toggle lock - push or pull elevating legrests Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 13
Swing back or removable, WEIGHT 36½ lbs space saver* or conventional, desk* or full length and in fixed* or adjustable height HANDRIMS Steel*, aluminum and plastic WEIGHT LIMIT 350 lbs (154 kg) coated Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 14
* These options are standard for this model. ** 9000 SL - The seat-to-floor heights are based on urethane tires. If wheelchair is equipped with pneumatic (properly inflated) or solid rubber tires, add ¼-inch to the measurements listed above. These heights can vary +/- ¼-inch due to tire wear and manufacturing tolerances.
NOTICE THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE. Check all parts for shipping damage and test before using. In case of damage, DO NOT use. Contact Invacare/Carrier for further instruction. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Seat-to-floor angle of 0° or 3°: If so equipped, anti-tippers MUST be attached at all times. Inasmuch as the anti-tippers are an option for 0° or 3° on this wheelchair (you may order with or without the anti-tippers), Invacare strongly recommends ordering the anti-tippers as a safeguard for the wheelchair user.
Invacare strongly recommends ordering the handrims as an additional safeguard for the wheelchair user. Invacare recommends that a non-folding device be installed to keep the wheelchair from being folded when left unoccupied in a public place.
Page 18
DO NOT attempt to recline or incline the wheelchair back without occupant. Back will raise abruptly and cause injury. Keep clear when reclining or inclining chair back to avoid injury. When learning a new assistance technique, have an experienced assistant help you before attempting it alone. Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 19
NEVER leave an unoccupied wheelchair on an incline. DO NOT attempt to stop the wheelchair while on a sloped surface. Repair and Service Information WARNING Unless otherwise noted, all service and adjustment should be performed while the wheelchair is unoccupied. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 20
ALWAYS wear your seat positioning strap. Inasmuch as the seat positioning strap is an option on this wheelchair (you may order with or without the seat positioning strap), Invacare strongly recommends ordering the seat positioning strap as an additional safeguard for the wheelchair user. The seat positioning strap is a positioning strap only. It is not designed for use as a safety device withstanding high stress loads such as auto or aircraft safety belts.
Page 21
N/A N/A N/A N/A N/A N/A N/A N/A N/A • CASTER SIZE CASTER POSITION N/A N/A • N/A N/A • WHEEL SIZE WHEEL POSITION N/A N/A • • N/A N/A USER CONDITION SEAT HEIGHT Part No. 1056953 Invacare® 9000 Series Wheelchair...
Weight Limitation WARNING The 9000 SL, 9000 XT, and 9000XT Recliner wheelchairs have a weight limitation of 250 lbs (114 kg). The 9000XDT wheelchair has a weight limitation of 350 lbs (159 kg). Invacare® 9000 Series Wheelchair...
WARNING Invacare does not recommend the use of its wheelchairs as a weight training apparatus. Invacare wheelchairs have not been designed or tested as a seat for any kind of weight training. If occupant uses said wheelchair as a weight training apparatus, Invacare shall not be liable for bodily injury or damage to the wheelchair and the warranty is void.
Use this information only as a basic guide. The techniques that are discussed on the following pages have been used successfully by many. Individual wheelchair users often develop skills to deal with daily living activities that may differ from those described in this manual. Invacare recognizes and encourages each individual to try what works best for him/her in overcoming architectural obstacles that they may encounter.
Page 25
Position wheelchair as close as possible to the desired object. Point front casters forward to create the longest possible wheelbase. Reach back only as far as your arm will extend without changing your sitting position. FIGURE 1 Backwards Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 26
Roll the wheelchair forward and slowly lower the front of the wheelchair in one continuous movement onto the sidewalk. Push the wheelchair forward until the rear wheels roll up and over the curb. Step Tube FIGURE 2 Method 1 - Wheelchair With Step Tubes Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 27
Extreme caution is advised when it is necessary to move an occupied or unoccupied wheelchair up or down the stairs. Invacare recommends that, if possible, the user be removed from the wheelchair prior to moving. Invacare recommends using two assistants and making thorough preparations. Make sure to use only secure, non-detachable parts for hand-held supports.
Page 28
The wheelchair should not be lowered until the last stair has been negotiated and the wheelchair has been rolled away from the stairway. If necessary, rotate the anti-tippers so the wheels are facing down. FIGURE 4 Stairways Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 29
Shift body weight into seat with transfer. During independent transfer, little or no seat platform will be beneath FIGURE 5 Transferring To and From Other Seats you. Use a transfer board if at all possible. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Bar on page 75. Engage both wheel locks, open the footrest/legrest for clearance FIGURE 6 Unfolding and Folding Wheelchair and transfer into the wheelchair. Refer to Transferring To and From Other Seats on page 29. Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 31
FIGURE 8 Folding Solid-Seat Models Continue to close the wheelchair by grasping the armrest furthest from you and pulling the armrest towards you. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Inspect seat positioning strap for any signs of wear. Ensure buckle latches. Verify hardware that attaches strap to frame is secure and undamaged. Replace if necessary. ❑ Inspect tires for flat spots and wear. ❑ Check pneumatic tires for proper inflation. Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 33
Ensure all spokes are uniformly tight. ❑ Inspect axle assembly for proper tension by spinning caster. Caster should come to a gradual stop. ❑ Inspect front caster for cracked, bent or broken spokes. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 34
Ensure that casters are free of debris. ❑ Ensure wheel bearings are clean and free of moisture. ❑ Clean upholstery and armrests. ❑ Check that all labels are present and legible. Replace if necessary. Invacare® 9000 Series Wheelchair Part No. 1056953...
Flutter and Rattles Chair Right/Left Check tires for correct and equal pressure Check for loose nuts and bolts. Check caster headtube angle. Check that rear wheels are equally spaced away from seat frame. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Check all parts for damage or wear and replace. Check all parts for proper adjustment. 9000 SL/9000 XT ONLY - Keep quick-release axles free of dirt and lint to ensure positive locking and proper operation. Refer to Adjusting Quick-Release Axle on page 58.
Page 37
Rubber Tires - 30% or more of the tire has worn away. Invacare recommends that tires and casters be replaced every five years. Periodically check handrims to ensure they are secured to the rear wheels. If loose, have them tightened by a qualified technician.
Page 38
4 SAFETY INSPECTION/TROUBLESHOOTING Replacing/Repairing Rear Wheel Tire/Tube WARNING Replacement of solid urethane tires is not recommended. If the solid urethane tire needs repaired, Invacare recommends replacing the complete wheel assembly. Replacement of rear wheel tube must be performed by a qualified technician.
To release the front rigging, push the front rigging release lever inward, rotate front rigging outward. Removing Push the front rigging release lever inward. Rotate swingaway front rigging assembly outward. Lift the swingaway front rigging assembly off the hinge pins. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 40
5 FRONT RIGGINGS Front Rigging Release Lever Front Rigging Hinge Pins Assembly Footplate Hinge Plates Swingaway footrest shown. FIGURE 1 Installing/Removing Front Riggings Invacare® 9000 Series Wheelchair Part No. 1056953...
Rotate cam lock lever down to locked position. FIGURE 2 Spring Button Height Repeat this procedure for the other footplate, if necessary. Reinstall the swingaway front rigging assembly. Refer to Installing/Removing Front Riggings on page 39. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 42
Using a screwdriver to hold the threaded rivet in position, securely tighten the button head screw. Torque to 32 in-lbs. Rotate cam lock lever down to locked position. 10. Repeat STEPS 1-9 to adjust the remaining footrest. Invacare® 9000 Series Wheelchair Part No. 1056953...
To lower the elevating legrest, perform the following: Support user leg with one hand. Push release lever downward with other hand. FIGURE 4 Raising/Lowering Elevating Legrest Assembly Part No. 1056953 Invacare® 9000 Series Wheelchair...
Remove the threaded rivet and button head screw securing the footplate assembly to the front rigging support. Remove the mounting screw, spacer and locknut that secure the heel loop to the footplate. Remove existing heel loop from slide tube. Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 45
Swingaway Spacer Adjustment Swingaway Assembly Assembly Front Rigging Holes Front Rigging Adjustment Hole Cam Lock Lever Footplate Footplate Cam Lock Assembly Assembly Lever Swingaway footrest shown. Locknut Footplate FIGURE 6 Replacing Heel Loop Part No. 1056953 Invacare® 9000 Series Wheelchair...
Lock the armrest by pressing the height adjustment lever into the down (vertical) position when the desired armrest height is Unlocked (Horizontal) achieved. FIGURE 1 Adjusting Armrest Height Repeat STEPS 1-3 for other armrest. Invacare® 9000 Series Wheelchair Part No. 1056953...
Armrest release lever MUST be in the unlocked Rear position when placing armrest into the arm sockets. Arm Socket Lock the swing-back arms by rotating the armrest release lever towards the inside of the wheelchair. FIGURE 2 Using Swing-Back Arms Part No. 1056953 Invacare® 9000 Series Wheelchair...
9000XDT: Use only heavy duty, fixed height armrests on 9000XDT wheelchairs. Otherwise injury or damage may occur. Make sure the armrest release lever is in the locked position before using the wheelchair. This procedure does not apply to 9000 SL wheelchairs equipped with space saver arms. Removing Armrests Armrest...
For space saver arms - position arm socket on inside of frame. Make sure arm sockets sit flush against the wheelchair frame. Reinstall the hex screw and locknut that secure the arm socket to the wheelchair frame and securely tighten. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 50
Wheelchair Frame) Arm Socket (Inside of Hex Screw Wheelchair Frame) Locknut Hex Screw Locknut Left Side of Wheelchair Frame Right Side of Wheelchair Frame Wheelchair Frame FIGURE 5 Converting Between Space Saver and Conventional Arms Invacare® 9000 Series Wheelchair Part No. 1056953...
16 inch seat depth has 8 mounting screws and 17-20 inch seat depths have 10 mounting screws. Remove the existing seat upholstery from the wheelchair frame. Install new seat upholstery by reversing STEPS 1-2. 16-inch Seat Depth Mounting Screws Seat Upholstery Wheelchair Frame FIGURE 1 Replacing Seat Upholstery Part No. 1056953 Invacare® 9000 Series Wheelchair...
Slip-On Back Upholstery This procedure applies to 9000 XT/XDT and 9000 SL with adjustable back angle option. Remove the two mounting screws and washers that secure the existing back upholstery to the back canes (Detail “A”). Invacare® 9000 Series Wheelchair...
Page 53
Secure the new back upholstery to the back canes with the existing mounting screws and washers. DETAIL “A” DETAIL “B” Back Upholstery Washer Back Cane Wheelchair Frame Mounting Spacers Screws Locknuts Mounting Screws Back Cane FIGURE 8 Replacing Back Upholstery - Slip-On Back Upholstery Part No. 1056953 Invacare® 9000 Series Wheelchair...
These adjustments must be performed by a qualified technician. Adjusting Seat Width WARNING Adjusting the seat width of the 9000SL, 9000XT, 9000XDT MUST be performed by a qualified technician. Invacare® 9000 Series Wheelchair Part No. 1056953...
NEVER use quick-release axles on the 9000 XT (models with offset axle option only), 9000XDT or 9000XT recliner wheelchairs. If replacing the same size rear wheel, note the mounting position on the wheelchair frame for proper reinstallation of the new rear wheel. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 56
The detent balls MUST be protruding past the inside of the rear wheel axle bushing for a positive lock (Detail "A"). Keep detent balls clean. Invacare recommends inserting quick-release axles with the detent pin to the inside of the wheelchair to prevent accidental release during contact activities.
Removing/Installing Rear Wheels on page 77. Remove the mounting screws that secure the handrim to rear wheel. Remove the existing handrim. FIGURE 2 Replacing Rear Wheel Handrim Install the new handrim by reversing the preceding steps. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Make sure detent pin and detent balls of the quick-release axles are fully released before operating the wheelchair. Keep detent balls clean. Reinstall rear wheel onto the wheelchair. Refer to Removing/Installing Rear Wheels on page 55. Invacare® 9000 Series Wheelchair Part No. 1056953...
9000XDT ONLY - Make sure to reinstall the 2-inch extension spacer. See Detail "A". Slide in the new six/eight-inch front caster and fork. Reassemble by reversing STEPS 1-3. Repeat STEPS 1-5 for the opposite front caster. Invacare® 9000 Series Wheelchair Part No. 1056953...
Caster Headtube Fork Stem Front Caster and Fork FIGURE 1 Installing/Replacing Six or Eight-Inch Front Casters and Forks Adjusting Front Caster Mounting WARNING Front Caster mounting adjustments MUST be performed by a qualified technician. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Seat-to-floor angle of 0° or 3°: If so equipped, anti-tippers MUST be attached at all times. Inasmuch as the anti-tippers are an option for 0° or 3° on this wheelchair (you may order with or without the anti-tippers), Invacare strongly recommends ordering the anti-tippers as a safeguard for the wheelchair user.
Page 63
Press the release buttons on the wheeled portion of the anti-tipper and slide it up or down to achieve the 1-1/2 to 2-inch clearance (Detail “A”). Check to make sure that the release buttons are fully engaged in adjustment holes. DETAIL “A” DETAIL “B” Anti-Tipper Height Release Buttons 2 inches 1½ inches Anti-Tipper Length FIGURE 2 Adjusting Anti-Tippers Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 64
9000 XT WITH 0° SEAT-TO-FLOOR ANGLE (ADULT/HEMI FRAME) CASTER SIZE SEAT-TO-FLOOR HEIGHT ANTI-TIPPER LENGTH/HEIGHT ON CHAIR OFF CHAIR KIT NUMBER 8-inch 17½, 18½ inches 13½ / 3½-inches 1058836 6-inch 16½, 17½-inches 13½ / 3½-inches 1058366 15½-inches 11-5/8 / 2-3/4-inches 1058752 Invacare® 9000 Series Wheelchair Part No. 1056953...
10 ANTI-TIPPERS/WHEEL LOCKS 9000 XT WITH 3° SEAT-TO-FLOOR ANGLE AND 9000 SL (ADULT/HEMI FRAMES) CASTER SIZE SEAT-TO-FLOOR HEIGHT ANTI-TIPPER LENGTH/HEIGHT ON CHAIR OFF CHAIR KIT NUMBER 8-inch 17½, 18½, 19½-inches 13½ / 3½-inches 1360 1058836 6-inch 15½, 16½, 17½-inches 11-5/8 / 2¾-inches...
Page 66
Reposition the wheel lock so that when engaged, the wheel lock shoe embeds the tire 1/8-inch (3/16-inch for pneumatic tires) and holds the occupied wheelchair in place when pushed. Securely tighten the bolt and locknut or socket screws securing the wheel lock to the wheelchair frame. Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 67
11. Engage both wheel locks and ensure the occupied wheelchair is held in place when pushed. WARNING If wheel locks DO NOT hold the occupied wheelchair in place, contact a qualified technician. Otherwise injury or damage may occur. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Securely tighten the two set screws to secure the wheel lock shoe extension to the wheel lock shoes. Set Screws Wheel Lock Shoe Extension Pointed End of Wheel Lock Shoe Extension Slot FIGURE 5 Installing Wheel Lock Shoe Extensions Part No. 1056953 Invacare® 9000 Series Wheelchair...
ALWAYS wear your seat positioning strap. Inasmuch as the seat positioning strap is an option on this wheelchair (you may order with or without the seat positioning strap), Invacare strongly recommends ordering the seat positioning strap as an additional safeguard for the wheelchair user. The seat positioning strap is a positioning strap only. It is not designed for use as a safety device withstanding high stress loads such as auto or aircraft safety belts.
Page 71
11. When the back reaches the desired angle, slowly let go of the release handles. 12. To return the back to the full upright position, reverse STEPS 8-11 keeping in mind proper body mechanics. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Unfasten straps of headrest pillow from behind headrest Upholstery Strips upholstery. Strap Secure straps of the new headrest pillow onto fastening strips located on back of headrest upholstery. Strap Headrest Pillow FIGURE 2 Replacing Headrest Pillow Invacare® 9000 Series Wheelchair Part No. 1056953...
Secure the new back upholstery to the back canes with the existing twelve mounting screws. Securely tighten. Install the knob removed in STEP 1 to secure spreader bar to side frame. Securely tighten both knobs. FIGURE 3 Replacing Headrest/Back Upholstery Part No. 1056953 Invacare® 9000 Series Wheelchair...
Adjusting the seat depth of the 9000XT Recliner MUST be performed by a qualified technician. 11.4 Adjusting Seat Width WARNING Adjusting the seat width of the 9000XT Recliner MUST be performed by a qualified technician. Invacare® 9000 Series Wheelchair Part No. 1056953...
Remove both locking knobs that secure the spreader bar to the side frames. Remove existing spreader bar. Side Frame Secure new spreader bar to the side frames with the existing locking knobs. Securely tighten. FIGURE 6 Replacing Spreader Bar Part No. 1056953 Invacare® 9000 Series Wheelchair...
Seat-to-floor heights are Anti-Rattle approximate depending on front caster type. The anti-tipper height measurement MUST be taken DETAIL “B” with the anti-tipper on a flat surface. Anti-Tipper Height Anti-Tipper Length FIGURE 7 Installing Anti-Tippers Invacare® 9000 Series Wheelchair Part No. 1056953...
DO NOT attempt to stop a moving wheelchair with wheel locks. Wheel locks are not brakes. Perform one of the following: • Engage - Apply pressure downward on wheel lock handle. • Disengage - Lift up on wheel lock handle. Invacare® 9000 Series Wheelchair Part No. 1056953...
Repeat the above procedures until the wheel lock holds the wheelchair. Repeat STEPS 1-5 for the opposite wheel lock. Ensure the wheel lock shoe completely covers the tire contact area. See Detail "A" for proper positioning of wheel lock shoe. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Pivot Screw Install the I. V. rod support tube. Refer to Installing I.V. Rod Support Screw (STEP 4) Tube on page 82. (STEP 6) Locknut (STEPS 3 and 5) FIGURE 12 Installing I.V. Holder Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 82
Adjust I.V. rod to desired height. Refer to Adjusting I.V. Rod on Rear page 82. Cylinder Adjusting I.V. Rod Mounting Screw Loosen adjustment knob. Adjust I.V. Rod to desired height. Securely tighten adjustment knob. FIGURE 13 Installing I.V. Rod Support Tube - Adjusting I.V. Invacare® 9000 Series Wheelchair Part No. 1056953...
Telescoping I.V. Rod Holder Only - Secure the top of the telescoping I.V. rod holder to the wheelchair frame with a tie wrap. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 84
The user and O2 cylinder MUST be removed from the wheelchair when cleaning the O2 holder. Otherwise, injury or damage may occur if cleaning agents come in contact with the O2 regulator. Remove the user, any I.V. medication and the O cylinder from the wheelchair. Invacare® 9000 Series Wheelchair Part No. 1056953...
Reinstall the back upholstery screw and washer and secure the strap to the wheelchair. Locknut The base cleans easily with any chrome or glass cleaner. The strap can be cleaned with mild soap and water. FIGURE 2 Installing Crutch and Cane Carrier Part No. 1056953 Invacare® 9000 Series Wheelchair...
AXLE MOUNTING POSITION (ON BRACKET) (ON WHEELCHAIR) ADULT DOWN RIGHT ADULT DOWN LEFT HEMI RIGHT BOTTOM HEMI LEFT BOTTOM * "A" and "B" are stamped on the sides of the amputee bracket. Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 87
Make sure the axle spacer is pointing towards the outside of the wheelchair. Install the two hex screws and locknuts that secure the amputee bracket to the wheelchair. Install the rear wheels onto the wheelchair. Refer to Removing/Installing Rear Wheels on page 55. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 88
Adult Right Adult Left Axle Spacer Toward Outside of Axle Spacer Toward Outside of Wheelchair Wheelchair Rear View of Amputee Bracket Hex Screws Hemi Left Hemi Right FIGURE 4 Installing Amputee Bracket - 9000SL/XT Invacare® 9000 Series Wheelchair Part No. 1056953...
Page 89
Make sure the axle spacer is pointing towards the outside of the wheelchair. Install the two hex screws and locknuts that secure the amputee bracket to the wheelchair. Install the rear wheels onto the wheelchair. Refer to Removing/Installing Rear Wheels on page 55. Part No. 1056953 Invacare® 9000 Series Wheelchair...
Page 90
Hemi Left Hemi Right Right Left Adult Right Adult Left Axle Spacer Toward Outside of Wheelchair Hex Screws Rear View of Amputee Bracket Tall Left Tall Right Locknuts FIGURE 5 Installing Amputee Bracket - 9000XDT Invacare® 9000 Series Wheelchair Part No. 1056953...