Download Print this page
Fluke 51 II Specifications
Fluke 51 II Specifications

Fluke 51 II Specifications

Fluke 50 series ii thermometers specifications
Hide thumbs Also See for 51 II:

Advertisement

50 Series II Thermometers
Fluke 54 II
Fluke 51 II
Fluke 52 II
Included Accessories
Impact absorbing holster
Two bead probe thermocouples 80PK-1
(54+52)
One bead probe thermocouple 80PK-1
(51+53)

Ordering Information

Fluke 51 II
Thermometer
Fluke 52 II
Thermometer
Fluke 53 II
Thermometer
Fluke 54 II
Thermometer
FVF-SC1
FlukeView Forms
Software including
interface cable
11080-eng Rev. 01
Laboratory accuracy. Wherever you go.
The Fluke 50 Series II contact
thermometers offer fast response and
laboratory accuracy (0.05% + 0.3°C) in
a rugged handheld test tool.
Large backlit dual display shows any
combination of T
T
-T
(52 and 54 only) plus MIN, MAX,
1
2
or AVG
Relative time clock on MIN, MAX, and
AVG provides a time reference for major
events
Electronic Offset function allows
compensation of thermocouple errors to
maximize overall accuracy
Readout in °C, °F, or Kelvin (K)
Sleep mode increases battery life
Battery door allows easy battery
replacement without breaking the
calibration seal
Thermocouple types
Number of inputs
Time stamp
Splash/Dust resistant
Dual display with backlight
Min/Max/Avg recording
(T
-T
) True differential
1
2
Data logging up to 500 points
IR data port for interface
to PC
Compatible with optional
FlukeView Software
Temperature range:
Fluke 53 II
J-type
Thermocouples
K-type
Thermocouples
T-type
Thermocouples
E-type
Thermocouples
N-type** Thermocouples
R** and S-type** Thermocouples
Temperature accuracy
Above –100°C (-148°F):
J, K, T, E, and N-type**
R** and S-type**
Below –100°C (-148°F):
J, K, E, and N-types
T-type
**Only the Fluke Models 53 and 54 Series II thermometers are capable of measuring N, R, and S-type thermocouples.
Battery life: 1000 hours typical, AA
Size (HxWxD): 173 x 86 x 38 mm
C25
Additional features for the 53 and
54 Series II:
Data Logging up to 500 points of data
with user adjustable recording interval
Real time clock captures the exact time of
, T
(52 and 54 only),
day when events occur
1
2
Recall function allows logged data to be
easily reviewed on the meter display
IR communication port allows data to be
exported to optional FlukeView
Temperature PC software
51 II
J,K,T,E
Single
Relative Time
Relative Time
-210°C to 1200°C
-200°C to 1372°C
-250°C to 400°C
-150°C to 1000°C
-200°C to 1300°C
0°C to 1767°C
± [ 0.05% + 0.3°C (0.5°F) ]
± [ 0.05% + 0.4°C (0.7°F) ]
± [ 0.20% + 0.3°C (0.5°F) ]
± [ 0.50% + 0.3°C (0.5°F) ]
Weight: 0.4 kg
Three Year Warranty
Recommended Accessories
80PK-26
80PK-25
®
Features
52 II
53 II
54 II
J,K,T,E
J,K,T,E,N,R,S
J,K,T,E,N,R,S
Dual
Single
Dual
Time of Day
Time of Day
Specifications
(-346°F to 2192°F)
(-328°F to 2501°F)
(-418°F to 752°F)
(-238°F to 1832°F)
(-328°F to 2372°F)
(32°F to 3212°F)
FVF-SC1
TPAK

Advertisement

loading

This manual is also suitable for:

53 ii52 ii50 series ii54 series ii