Model UT60A: OPERATING MANUAL
Measurement Operation (2)
Note
l
In each range, the Meter has an input impedance
of 10M . This loading effect can cause measurement
errors in high impedance circuits. If the circuit
impedance is less than or equal to 10k , the error
is negligible (0.1% or less).
l
When AC voltage measurement has been
completed, disconnect the connection between the
testing leads and the circuit under test.
DC Voltage Measurement
Hz
V
The DC Voltage ranges are: 400.0mV, 4.000V, 40.00V,
400.0V and 1000V. To measure DC voltage, connect
the Meter as follows:
1.
Insert the red test lead into the HzV
the black test lead into the COM terminal.
2.
Set the rotary switch to Hz
is default or press BLUE button to select DC
measurement mode.
3.
Connect the test leads across with the object being
measured.
The measured value shows on the display.
(see figure 4)
Select
red
black
(figure 4)
V; DC measurement
20
terminal and