ReliOn® Confirm/micro test strips
These test strips need only 0.3 µL (microliter) of blood. Apply blood to the
sample application tip. The strip draws in the blood by capillary action.
The blood glucose reacts with the enzyme on the test strip. This produces
a current proportional to the blood glucose level. The meter detects the
current and converts it into a blood glucose reading.
Test Strip Bottle
Front
Code No.
Code No.
(F-2, 3 or 4)
(F-2, 3 or 4)
Expiration date
F-4
Back
Test Strip
F- 4
Insert this
way into
10
meter
Contact Bars
Sample
Application
Tip (Apply blood
to end of test strip.
Capillary draw will
pull blood to
reaction site).