Download Print this page
Quincy lab Analog Series Operating Manual

Quincy lab Analog Series Operating Manual

Lab oven

Advertisement

Quick Links

uincy
ab
SPECIFICATIONS
(-1) indicates 230V model
INTERIOR DIMENSIONS
EXTERIOR DIMENSIONS
TEMPERATURE RANGE
CONTROL STABILITY
STANDARD ELECTRICAL
(230V) Model
( 230V) Model
(230V) Model
Common Unit Construc�on
Exterior: Powder Coated Steel
Insulation: Fiberglass
Thermo-control: Bimetal(10 series) / Hydraulic
Issue Rev 10.22
Copyright Quincy Lab Inc. 2022
Analog Series Air Forced (AF) Lab Oven
MODEL
10AF (-1)
12x8.25x10
INCHES W x H x D
30.5x21x25.4
(CM) W x H x D
14x20.5x12.25
INCHES W x H x D
35.5x51x31.7
(CM) W x H x D
450° / 232°
6° / 3°
VOLTS / AMPS
120 / 6.3
VOLTS / AMPS
230 / 3.5
WATTS
WATTS
FREQUENCY
50/60 Hz
PLUG / NEMA
5-15P
6-15P
PLUG / NEMA
WEIGHT (lbs)
SHIPPING
STAND ALONE
Quincy Lab Inc. 109 Shore Dr.
MODEL
20AF (-1)
13x11x13
33x28x33
15x25x15
38x63.5x38
450° / 232°
6° / 3°
120 / 7.9
230 / 4.3
750
950
800
1000
50/60 Hz
5-15P
6-15P
50
66
44
49
Interior: Aluminized Steel
Motor:
Thermally Protected
Heater: Resistive-Tubular Incoloy
Burr Ridge, Illinois 60527
Operating Manual
For models 10AF, 20AF, 30AF, & 40AF
10AF-1, 20AF-1, 30AF-1, & 40AF-1
Carton Contents
(1) AF Series Oven
(2) Adjustable nickle-plated wire shelves
(4) Shelf brackets
Approvals:
Underwriter's UL / CUL, United States
Canadian for laboratory equipment.
Compliance: UL Standard 61010-1
IEC 61010-1.
NOT FOR USE WITH FLAMMABLE
LIQUIDS OR GASES
MODEL
MODEL
30AF (-1)
40AF (-1)
18x14.13x12
18x19.63x14
45.7x35.5x30.5
45.7x50x35.5
20x29x14
20x31.5x16
51x73.7x35.5
51x80x40.7
450° / 232°
450° / 232°
6° / 3°
6° / 3°
120 / 12.5
120 / 12.5
230 / 7.0
230 / 7.0
1500
1500
1600
1600
50/60 Hz
50/60 Hz
5-15P
5-15P
6-15P
6-15P
82
112
74
95
1800-482-HEAT (4328)
PAGE 1

Advertisement

loading

Summary of Contents for Quincy lab Analog Series

  • Page 1 Analog Series Air Forced (AF) Lab Oven uincy Operating Manual For models 10AF, 20AF, 30AF, & 40AF 10AF-1, 20AF-1, 30AF-1, & 40AF-1 Carton Contents (1) AF Series Oven (2) Adjustable nickle-plated wire shelves (4) Shelf brackets Approvals: Underwriter’s UL / CUL, United States Canadian for laboratory equipment.
  • Page 2 WARNING IMPORTANT WARNING Refers to an important note in the usage of the unit Warns of a possible electrical shock Warns of a possible risk of fire CAUTION WARNING Cautions of HOT exterior surface during operation Warns of possible injury or muscle strains, use assistance when moving or lifting Safety Precautions Read Operating Instructions thoroughly prior to operation Operate the oven in accordance with all state, local, and federal laws that may govern its usage for your specific...
  • Page 3 General Operation (continued) For 10AF models: Rotate the dial clockwise to around the number '8'. The “heat/cycle” light will come on indicating heater activity. Monitor the reading on the door thermometer until it reaches the desired temperature. Once the desired temperature is reached, rotate the control knob very slowly counter-clockwise until the heat-cycle light turns off. The thermostat will then cycle automatically maintaining the set temperature.
  • Page 4: Troubleshooting

    Burr Ridge, Illinois 60527 Limited Warranty Quincy Lab, Inc. warrants to the original purchaser that this product will be free from defects in material and workmanship under normal use throughout the warranty period. The standard warranty period for this instru- ment is twenty four (24) months from date of shipment.
  • Page 5 LAB OVEN SERIES Calibra�on and Temperature Range Adjustement for Analog Lab Ovens A calibration to the knob can be performed if the chamber temperature is off by 15 degrees Fahrenheit or more from the temperature number printed on the dial on models 20,30,and 40 GC/AF. A temperature range adjustment can be performed to the oven when not able to reach lower or higher temperatures when set with the dial.

This manual is also suitable for:

Analog 10afAnalog 10af-1Analog 20afAnalog 20af-1Analog 30afAnalog 30af-1 ... Show all