  • Page 2 European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
  • Page 3: Safety Information

    This instrument is a 4½ digit, compact-sized portable digital thermometer designed to use external K/J/T/E/R/S/N-type thermocouples as temperature sensor. Temperature indication follows Reference Temperature/Voltage Tables (N.I.S.T. Monograph 175 Revised to ITS-90) for K/J/T/E/R/S/N-type thermo- couples. One K-type thermocouple are supplied with the thermometer.
  • Page 4: Specifications

    ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable Measurement Range: Thermocouple Range K-TYPE -200°C to 1372°C, -328°F to 2501°F J-TYPE -210°C to 1050°C, -346°F to 1922°F T-TYPE -200°C to 400°C, -328°F to 752°F E-TYPE -210°C to 790°C, -346°F to 1454°F R-TYPE S-TYPE N-TYPE Resolution:...
  • Page 5 Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error. K/J/T/E-TYPE ±(0.05% rdg + 0.3°C) -50°C to 1370°C ±(0.05% rdg + 0.7°C) -50°C to -210°C ±(0.05% rdg + 0.6°F) -58°F to 2498°F ±(0.05% rdg + 1.4°F) -58°F to -346°F N-TYPE ±(0.05% rdg + 0.7°C) -50°C to 0°C...
  • Page 6 ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 9999.9. Overload: "----.-" is display. Battery: Standard 9V battery. Battery Life: 100 hours typical with carbon zinc battery.
  • Page 7: Operating Instructions

    OPERATING INSTRUCTIONS Power Switch key turns the thermometer on or off. In the MIN MAX record mode can not power off, must leave MIN MAX record mode then power off. °C/°F Selecting the Temperature Scale Readings are dual displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F).
  • Page 8 The TYPE key circulating selects the K/J/T/E/R/S/N type thermocouple as input. When the thermometer is turned on, it is set to the type selected that was in use when the thermometer was last turned off. MIN MAX with Time record Mode (only Main display)
  • Page 9 Push MIN MAX key to cycle through the Max/Min/Avg/Present readings. If an overload is recorded, the averaging function is stopped and average value display"----.-". The true average of all the reading taken over at least 22 hours preiod can be displayed. If 22 hours is exceeded, new averages are no longer calculated.
  • Page 10 Hi/Lo LIMITS mode (only Main display) Press Hi/Lo LIMITS key to enter the Hi/Lo Limits comparative mode, "LIMIT" is displayed. In the this mode the automatic power-off feature is disabled and REL key, HOLD key, MAX/MIN key are disabled, when input temperature value exceed Hi.
  • Page 11 3. Time set mode, (Press ENTER key can escape Time set mode) = = : = = = = is displayed in third display. Time (hours, minutes, seconds) value is entered via overlay numbers, then press overlay ENTER key. Time start from set time value and exit set mode.
  • Page 12: Operator Maintenance

    OPERATOR MAINTENANCE WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery cover.
  • Page 15 ° °...
  • Page 16: Output Data Formats

    OUTPUT DATA FORMATS The data format consists of 32 bytes. byte T1 present status S: Negative sign or positive sign H: Hexadecimal digits D: Numeric digits A: Capital Letter or "-" sign CR: Carriage return character nl: newline character T1 polarity 02 —...
  • Page 17 17 — MSD Hours 18 — LSD 19 — MSD Minutes 20 — LSD 21 — MSD Seconds 22 — LSD 23 — Record mode 24 — REL mode 25 — HOLD mode 26 — LIMITS mode 27 — Hi LIMITS 28 —...
  • Page 18 507R 1. Install the RSR232C cable between computer "COM1" and Thermometer RS232C port. 2. Turn on Thermometer. 3. Utilize the Quick BASIC for inputting programs. 4. Use the capital control letter and input "A" letter from the keyboard to start sending messages to computer.
  • Page 19 HH 506(R)/HH 507(R) CALIBRATION PROCEDURE Note: The following calibration procedure should perform only by qualified technicians who have access to the items as following. Equipment: The class of calibrator had better 10 times greater than the measured meter. 1. Set the jumper position on J1 position. 2.
  • Page 20 OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
  • Page 21: Data Acquisition