BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
INSTRUCTION MANUAL
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes
before taking a measurement. Rest for at least 15 minutes before taking the
measurement.
5 series
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
Blood Pressure Monitor
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your left arm.
Model BP742
5. Sit on a chair with your feet flat on the floor. Rest your left arm on table so that the
cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return
to the condition prior to taking the blood pressure measurement. You may need to
increase the wait time depending on your individual physiological characteristics.
8. Keep a record of your blood pressure and pulse readings for your physician. A
INTRODUCTION
single measurement does not provide an accurate indication of your true blood
pressure. You need to take and record several readings over a period of time. Try to
®
Thank you for purchasing the OMRON
BP742 5 series Automatic Blood Pressure
measure your blood pressure at the same time each day for consistency.
Monitor.
Fill in for future reference.
KNOW YOUR UNIT
DATE PURCHASED: __________________
Main Unit:
Display
SERIAL NUMBER: ____________________
• Staple your purchase receipt here
Air Jack
• Register your product on-line at
www.register-omron.com
Date/Time setting (
)
Your new blood pressure monitor uses the oscillometric method of blood pressure
button
measurement. This means the monitor detects your blood's movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
Components:
monitor does not need a stethoscope so the monitor is simple to use.
The BP742 comes with the following components:
• Monitor • Wide Range Cuff • 4 "AA" Alkaline Batteries
• Instruction Manual printed in English and Spanish • Quick Start Guide
The BP742 5 series Automatic Blood Pressure Monitor is intended for home use.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
Wide Range Cuff
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
UNIT DISPLAY
SAVE THESE INSTRUCTIONS
Systolic blood pressure
Diastolic blood pressure
SAFETY INFORMATION
Memory symbol
(Displayed when viewing
INTENDED USE
values stored in memory)
This product is designed to measure the blood pressure and pulse rate of people within
Average Value symbol
the range of the designated arm cuff, following the instructions in this instruction
Movement error symbol
manual. It is mainly designed for general household use. Please read the "Safety
Irregular Heartbeat symbol
Information" in this instruction manual before using the unit.
Date/Time Display
To assure the correct use of the product, basic safety measures should always be
followed including the warnings and cautions listed in this instruction manual.
DISPLAY SYMBOLS
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
Indicates a potentially hazardous situation which, if not avoided,
IRREGULAR HEARTBEAT SYMBOL (
WARNING
could result in death or serious injury.
When the monitor detects an irregular rhythm two or more times during the
Indicates a potentially hazardous situation which, if not avoided,
measurement, the Irregular Heartbeat Symbol (
may result in minor or moderate injury to the user or patient or
CAUTION
measurement values.
damage to the equipment or other property.
An irregular heartbeat rhythm is defined as
a rhythm that varies by less than 25% from
OPERATING THE DEVICE
the average rhythm or 25% more than the
Contact your physician for specific information about your blood pressure.
average rhythm detected while the monitor
Self-diagnosis and treatment using measured results may be dangerous. Follow the
is measuring the systolic and diastolic
instructions of your physician or licensed healthcare provider.
blood pressure.
DO NOT adjust medication based on measurement results from this blood pressure
If the irregular heartbeat symbol (
monitor. Take medication as prescribed by your physician. Only a physician is
displays with your measurement results,
qualified to diagnose and treat High Blood Pressure.
we recommend you consult your
The monitor is not intended to be a diagnostic device.
physician. Follow the directions of your
doctor.
If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Contact a physician immediately.
MOVEMENT ERROR SYMBOL (
Read all of the information in the instruction manual and any other literature in the
The Movement Error Symbol (
box before operating the unit.
measurement. Please remove the arm cuff, and wait 2-3 minutes. Reapply the arm cuff
Do not use this device on infants or persons who cannot express their intentions.
and take another measurement.
AVERAGE VALUE SYMBOL (
Read the Special Conditions section of the instruction manual if your systolic
pressure is known to be more than 220 mmHg. Inflating to a higher pressure than
The Average Value Symbol (
necessary may result in bruising where the cuff is applied.
The most recent average reading appears on the display screen.
Operate the device only as intended. Do not use the device for any other purpose.
HEARTBEAT SYMBOL/HYPERTENSION INDICATOR (
Dispose of the device, components and optional accessories according to applicable
TAKING A MEASUREMENT
local regulations. Unlawful disposal may cause environmental pollution.
The Heartbeat Symbol (
) flashes on the display during the measurement. The
Heartbeat Symbol (
) flashes at every heartbeat.
Do not use a cellular phone near the device. It may result in an operational failure.
When the measurement is complete, the Heartbeat Symbol (
Use only Omron authorized parts and accessories. Parts and accessories not
display with your blood pressure and pulse rate if the reading is above 135 for the
approved for use with the device may damage the unit.
Systolic Blood Pressure value and/or above 85 for the Diastolic Blood Pressure
Do not use other types of batteries. This may damage the unit.
value.
Do not use in a location with moisture, or a location where water may splash on the
USING THE MEMORY FUNCTION
unit.
The Heartbeat Symbol (
) flashes when the blood pressure reading stored in the
memory is above 135 for the Systolic Blood Pressure value and/or above 85 for the
CARE AND MAINTENANCE
Diastolic Blood Pressure value.
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not submerge the device or any of the components in water.
Internationally recognized guidelines for upper limit of normal home blood pressure
readings are as follows:
Store the device and the components in a clean, safe location.
Changes or modification not approved by the manufacturer will void the user
Home Guideline for Upper Limit of Normal BP
warranty. Do not disassemble or attempt to repair the unit or components.
Systolic Blood Pressure
Diastolic Blood Pressure
This is a general guideline because blood pressure depends on age and morbidity.
A lower home blood pressure goal may be recommended for certain patients,
including diabetic patients, pregnant women, and patients with renal failure.
OMRON HEALTHCARE recommends you contact your physician if the Heartbeat
Symbol (
) flashes on the display with your measurement values.
BATTERY INSTALLATION
1. Press the
indicator on the battery cover and
slide the cover off in the direction of the arrow.
2. Install 4 "AA" size batteries so the + (positive) and
- (negative) polarities match the polarities of
the battery compartment as indicated.
3. Replace the battery cover.
BATTERY REPLACEMENT
Low Battery Indicator
When the Low Battery Indicator appears on the display, turn the monitor off and
remove all the batteries. Replace with four new batteries at the same time. Long-life
alkaline batteries are recommended.
Battery compartment
WARNING
If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.
START/STOP button
NOTES: • Turn the unit off before replacing the batteries. If the batteries are removed
USER ID
while the unit is still on, the date and time will be reset to that of the
previous use. However, the measurement results are not deleted.
Selection Switch
AC adapter Jack
• When the batteries are replaced, you may need to reset the date and time. If
Up/Down (
) buttons
"-:--" appears on the display, refer to "Setting the Date and Time".
• The supplied batteries may have a shorter life.
CAUTION
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
INSTRUCTION MANUAL
Air tube
5 series
Blood Pressure Monitor
Model BP742
5329898-7C
5 Series Arm Blood Pressure
INTRODUCTION
Monitor
Thank you for purchasing the OMRON
®
BP742 5 series Automatic Blood Pressure
Monitor.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
BP742
Air plug
Staple your purchase receipt here
pressure measurement. This means the monitor detects your blood's movement through
Your new digital blood pressure monitor uses the oscillometric method of blood
Instruction
your brachial artery and converts the movements into a digital reading. An
Quick Start
oscillometric monitor does not need a stethoscope so the monitor is simple to use.
The BP742 comes with the following components:
• Monitor • Standard Adult Arm Cuff
• Instruction Manual printed in English and Spanish • Quick Start Guide
home use.
The BP742 IntelliSense
Automatic Inflation Blood Pressure Monitor is intended for
®
Manual
SAVE THESE INSTRUCTIONS
Guide
Heartbeat symbol /
SETTING THE DATE AND TIME
Hypertension Indicator
Set the monitor to the current date and time before taking a measurement for the first
Pulse display
time. If the date and time are not set correctly, the measurement values stored in the
Deflation symbol
memory and the average values may not be correct.
User ID symbol (A/B)
Battery low indicator
1. TO START THE SETTING
Press the Date/Time Setting button (
) to adjust the date and time.
)
) will appear on the display with the
Normal Heartbeat
The year flashes on the display.
Pulse
Blood
pressure
Irregular Heartbeat
)
Short Long
Pulse
2. SETTING THE YEAR
Blood
The year can be set between 2010 and 2040. When the
pressure
display reaches 2040, it will return to 2010.
)
Press the Up button (
) to advance by one year.
Press the Date/Time Setting button (
) to set the current
) is displayed if you move your body during the
year.
)
The month flashes on the display.
) is displayed when you press the memory button.
NOTES: • Press the Down button (
) to decrease the
date and time values.
)
• Press and hold the button to increase or
decrease the date and time values faster.
3. SETTING THE MONTH
) flashes on the
Press the Up button (
) to advance by one month.
Press the Date/Time Setting button (
) to set the current
month.
The day flashes on the display.
4. SETTING THE DAY
Press the Up button (
) to advance by one day.
Press the Date/Time Setting button (
) to set the current
day.
The hour flashes on the display.
5. SETTING THE HOUR
135 mmHg
The time is set using AM or PM.
Press the Up button (
) to advance by one hour.
85 mmHg
Press the Date/Time Setting button (
) to set the current
hour.
The minute flashes on the display.
6. SETTING THE MINUTE
Press the Up button (
) to advance by one minute.
Press the Date/Time Setting button (
) to set the minute setting.
7. TO TURN THE MONITOR OFF
Press the START/STOP button.
NOTE: The monitor will automatically turn off five minutes after the minute has
been set.
SPECIAL CONDITIONS
APPLYING THE ARM CUFF
CAUTION
Inflating to a higher pressure than necessary may result in bruising where
CAUTION
the cuff is applied.
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
If your systolic pressure is known to be more than 220 mmHg, press and
CAUTION
hold the START/STOP button until the monitor inflates 30 to 40 mmHg
This device is intended for use in measuring blood pressure and
higher than your expected systolic pressure.
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
APPLYING THE CUFF ON THE LEFT ARM
1. Make sure the air plug is securely inserted
in the main unit.
NOTES: • Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
2. Remove tight-fitting clothing from your upper
arm.
3. Apply the cuff to your left upper arm so
Left Arm
1/2 inch (1 - 2 cm)
the arrow is centered on the inside of your arm
and aligned with your middle finger.
The air tube runs down the inside of your arm.
The bottom of the cuff should be
Arrow
Air Tube
approximately 1/2 " above your elbow.
4. Wrap the cuff firmly in place around your
arm using the cloth fastening.
APPLYING THE CUFF ON THE RIGHT ARM
USING THE MEMORY FUNCTION
When taking a measurement using the right arm, use this instruction for Step 3.
Right Arm
The monitor is designed to store the blood pressure and the pulse rate in the memory
Apply the cuff to your right upper arm so
for two people (USER A and USER B) every time a measurement is completed.
the arrow and the air tube runs along the inside
The monitor automatically stores up to 30 sets of measurement values (blood pressure
of your arm. The bottom of the cuff should be
and pulse rate) for each user (A and B). When 30 sets of measurement values are
approximately 1/2 " above your elbow.
stored, the oldest record is deleted to save the most recent values.
Arrow
NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air
AVERAGING FUNCTION
to the cuff.
The monitor calculates the average reading based on the three most recent sets of
measurement values taken within 10 minutes of the most recent reading.
NOTE: If two sets of measurement values are stored in the memory for the 10
minutes period, the average is based on the two sets of measurement values.
If one set of measurement values is stored, this is displayed as the average.
TAKING A MEASUREMENT
TO DISPLAY THE MEASUREMENT VALUES
1. Press the START/STOP button to turn the monitor on.
1. Press the START/STOP button to turn the monitor on.
2. Select your USER ID (A or B).
All symbols appear on the display.
3. Press the Up or Down button (
on the display when the average reading is shown.
4. Press the Up or Down button (
Press the Down button (
Average value → most recent set of measurement values → the older readings →
back to the average value
Press the Up button (
) repeatedly
2. Select your USER ID (A or B).
Average value → oldest set of measurement values → the more recent readings →
back to the average value
NOTES: • The measurement value number is displayed
briefly in the Pulse Display area when you press
the Up/Down buttons (
• The date and time are alternately displayed with
the measurement values.
3. Sit on a chair with your feet flat on the floor.
• If the date and time are not set, "-:--" is displayed
Place your arm on a table so the cuff is level
with the measurement values.
with your heart.
• Press the button repeatedly to display the next
Keep still and do not talk during measurement.
values.
4. Press the START/STOP button.
• Press and hold the button to display the values
The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically
faster.
determines your ideal inflation level.
Remain still and do not move your arm until the entire measurement process is completed.
TO DELETE ALL VALUES STORED IN THE MEMORY
The values stored in the memory are deleted by USER ID.
You cannot partially delete values stored in the memory. All values for
the user you select will be deleted.
1. Press the START/STOP button to turn the monitor on.
2. Select your USER ID (A or B).
3. While holding the Down button (
NOTE: To stop the inflation or measurement, push the START/STOP button. The
more than five seconds until the display appears as it does in the following
monitor will stop inflating, start deflating, and will turn off.
illustration.
5. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the display and the Heartbeat
Symbol (
) flashes.
4. Press the START/STOP button to turn the monitor off.
6. When the measurement is complete, the arm cuff completely
NOTE: The monitor will automatically turn off after five minutes.
deflates. Your blood pressure and pulse rate are displayed.
7. Press the START/STOP button to turn the monitor off.
NOTE: The monitor will automatically turn off after
five minutes.
WARNING
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician or licensed healthcare provider.
WARNING
The monitor is not intended to be a diagnostic device.
CARE AND MAINTENANCE
To keep your blood pressure monitor in the best condition and protect the unit from
damage, follow the directions listed below:
Do not forcefully bend the arm cuff or air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth.
Do not use any abrasive or volatile cleaners.
Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water. Do
not subject the monitor to extreme hot or cold temperatures,
humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Remove the batteries if the unit will not be used for three months or longer. Always
replace all the batteries with new ones at the same time.
Use the unit in a manner consistent with the instructions provided in this manual.
CAUTION
Use only Omron authorized parts and accessories. Parts and accessories
not approved for use with the device may damage the unit.
CAUTION
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
ERROR INDICATORS
SYMBOL
CAUSE
CORRECTION
Take another measurement and
Monitor could not
remain still until the measurement
detect pulse wave.
is complete.
Cuff under-inflated.
Cuff not applied
Remove the arm cuff.
correctly.
Read "Taking a Measurement".
Wait 2-3 minutes.
Cuff over-inflated.
Take another measurement.
Device error.
Contact Customer Service.
). The Average Value Symbol (
) appears
Movement during
Keep still and take another
) to display the sets of measurement values.
measurement.
measurement.
) repeatedly
Remove the arm cuff. Wait 2-3
minutes and then take another
Irregular or weak
measurement. Repeat the steps in
pulses are detected.
section "Applying the Arm Cuff".
If this error continues to appear,
contact your doctor.
Replace the four batteries.
Batteries are low.
).
Refer to "Battery Installation".
Alternating
display
TROUBLESHOOTING TIPS
PROBLEM
CAUSES AND SOLUTIONS
No power.
Replace all four batteries with new ones.
No display appears on the unit.
Check the battery installation for proper placement
of the battery polarities.
Blood pressure varies constantly. Many factors
), press the START/STOP button down for
including stress, time of day, how you wrap the
Measurement values appear too
cuff, may affect your blood pressure.
high or too low.
Review the sections "Before Taking a
Measurement" and "Taking a Measurement".
FCC STATEMENT
Note:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference
in a residential installation. The product generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If the product does
cause harmful interference to radio or television reception, which can be determined by
turning the product on and off, the user is encouraged to try to correct the interference
by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise emissions
from digital apparatus as set out in the interference-causing equipment standard entitled
"Digital Apparatus", ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux
appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:
"Appareils Numériques", ICES-003 édictée par le minister des communications.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment.
LIMITED WARRANTY
Your BP742 5 series Automatic Blood Pressure Monitor, excluding the arm cuff and
batteries, is warranted to be free from defects in materials and workmanship appearing
within 2 years from the date of purchase, when used in accordance with the instructions
provided with the monitor. The arm cuff is warranted to be free from defects in
materials and workmanship appearing within one year from the date of purchase when
the monitor is used in accordance with the instructions provided with the monitor. The
above warranties extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff
covered by the above warranties. Repair or replacement is our only responsibility and
your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the
address of the repair location and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to prevent
damage in transit. Because of possible loss in transit, we recommend insuring the
product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE
IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD
OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights
that vary by jurisdiction. Because of special local requirements, some of the above
limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Visit our web site at:
www.omronhealthcare.com
Call toll free:
1-800-634-4350
SPECIFICATIONS
BP742 (HEM-7200-Z)
Model:
LCD Digital Display
Display:
Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Measurement Range:
Pressure: ±3mmHg or 2% of reading
Accuracy/Calibration:
Pulse: ±5% of reading
Automatic by electric pump
Inflation:
Automatic pressure release valve
Deflation:
Automatic exhaust valve
Rapid Air Release:
Oscillometric method
Measurement Method:
1.5V 4 "AA" batteries
Power Source:
Approx. 300 uses with 4 new alkaline batteries
Battery Life:
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
Operating Temperature
/Humidity:
Storage Temperature
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
/Humidity/Air Pressure:
700-1060 hPa
Main Unit Weight:
Approximately 12 oz. (340g)
not including batteries
Approx. 5 l/2"(l) × 4 7/8"(w) × 3 3/8" (h)
Main Unit Dimensions:
(141 mm × 123 mm × 85 mm)
Approx. 6"(w) × 22 1/8"(l) (151 mm × 563 mm)
Cuff Size:
Cuff tube 35 3/8" (900 mm)
Fits arm circumferences 9" to 17"
Cuff Circumference:
(220 mm to 420 mm)
Up to 30 readings per user
Memory:
Main Unit, Wide Range Cuff, 4 "AA" Alkaline Batteries,
Contents:
Instruction Manual, and Quick Start Guide
Optional Accessories:
AC Adapter
NOTE: These specifications are subject to change without notice.
Manufactured for OMRON HEALTHCARE Co., Ltd.
Distributed by OMRON HEALTHCARE, INC.
1925 West Field Court
Lake Forest, IL 60045 U.S.A.
www.omronhealthcare.com
© 2012 OMRON HEALTHCARE, INC.
Made in Vietnam

Advertisement

loading

Summary of Contents for Omron BP742

  • Page 1 When taking a measurement using the right arm, use this instruction for Step 3. applicable local regulations. Unlawful disposal may cause environmental The BP742 comes with the following components: Your BP742 5 series Automatic Blood Pressure Monitor, excluding the arm cuff and pollution. Right Arm...
  • Page 2 Lea este manual de instrucciones detenidamente antes de usar la unidad. de movimiento El monitor ha sido diseñado para almacenar la presión arterial y el pulso en la memoria Su monitor de presión arterial automático BP742 serie 5, excluyendo el brazalete y las pilas, está SÍMBOLO CAUSA CORRECCIÓN...